5.5mm Diameter x 500mm CoCr Straight Rod, Model number… recallZ-1884-2017 · Xtant Medical Inc · initiated Apr 12, 2017
Class II Terminated All recalls posted in April 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Xtant Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The material type on the label may incorrectly state Ti6Al-4V ELI. The rods are composed of Cobalt Chromium.
FDA root cause category: Labeling mix-ups
What the company is doing
XTANT sent an Urgent Medical Device Removal letter dated 4/12/2017 to customers with a Return Response Acknowledgement and Receipt form. The letter identified the affected product, problem and actions to be taken. For questions call 406-570-7028.
Recall details
- Recall number
- Z-1884-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 12, 2017
- Posted by the FDA
- Apr 27, 2017
- Recalling firm
- Xtant Medical Inc
- Quantity
- 97 units
- Product code
- MNH
- FDA submissions
- K120832
- Distribution
- Worldwide distribution - US Nationwide in the states of: AZ, CA, CT, GA, LA, NV, OH, TX and in the country of Mexico.
Product description
5.5mm Diameter x 500mm CoCr Straight Rod, Model number X077-0420. The product is packaged in plastic bags with one piece per bag. Product Usage: The 5.5mm dia. X 500 mm CoCr Straight Rod is an accessory to the Fortex Pedicle Screw System. It is a straight rod connected between two or more pedicle screws to create a pedicle screw construct. Indications for Use: The Fortex Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. In addition, the system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
Code information (models, lots, serial numbers)
Part number X077-0420, lot number 041218
Other devices under the same FDA submission
5 devices cleared under K120832 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Fortex Pedicle Screw System N60001073Xtant Medical Holdings, Inc. · Orthosis, spondylolisthesis spinal fixation2 reports
- Fortex Pedicle Screw System X022-0042Xtant Medical Holdings, Inc. · Orthosis, spondylolisthesis spinal fixation2 reports
- Fortex Pedicle Screw System X022-0096Xtant Medical Holdings, Inc. · Orthosis, spondylolisthesis spinal fixation2 reports
- Fortex Pedicle Screw System X070-0011Xtant Medical Holdings, Inc. · Orthosis, spondylolisthesis spinal fixation2 reports
- Fortex Pedicle Screw System X022-0095Xtant Medical Holdings, Inc. · Orthosis, spondylolisthesis spinal fixation1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page