GIZA Vertebral Body Replacement recallZ-2072-2017 · Eden Spine Europe SA · initiated Mar 17, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Eden Spine Europe SA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Complaints that implants have been disassembled by surgeons because of unscrewing completely the locking screw.
FDA root cause category: Labeling design
What the company is doing
A correction letter to the users was sent on March 28, 2017 Stating the following: Eden Spine Europe has become aware of extremely rare incidents of disassembly of Giza devices, which is distributed by Stryker Spine in the USA. These incidents have occurred in limited situations when a surgeon has needed to unscrew the device in order to readjust the distraction. To mitigate this risk, a more precise description of the unlocking procedure in the Giza Surgical Technique was added to prevent surgeons from fully unscrewing the locking screw. POTENTIAL HAZARD In the unlikely event the Giza device was to disassemble, the potential hazard is that the surgeon may decide either to reassemble the device or change the implant to a new device. This could potentially lead to a short surgical time delay of 2 to 10 minutes. RISK MITIGATIONS Eden Spine Europe and Stryker have added the following information in the Giza Surgical Technique to mitigate the risk of disassembly of the device: Note: If adjustment is needed, the implant locking mechanism must be unscrewed 3.5 to 4 rotations from the tightened position, to free the superior cylinder from the inferior one. No more than 4 counterclockwise rotations should be applied to the locking screw, as doing so could potentially lead to disassembly of the implant's locking mechanism. The GIZATM Vertebral Body Replacement surgical technique is available on Stryke Spine's website. ACKNOWLEDGEMENT OF THIS NOTICE Please acknowledge within 5 days of receipt of this notice: - Fill out and send the following online for: https://fr.surveymonkey.com/r/5XYT3ZQ - Or Send an Email to the following address: [email protected] stating that this notice was received. For any questions you may contact: Christophe Zuanon, Regulatory Affairs & Quality Assurance Manager [email protected] Manager, [email protected]
Recall details
- Recall number
- Z-2072-2017
- Status
- Terminated
- Initiated by the company
- Mar 17, 2017
- Recalling firm
- Eden Spine Europe SA
- Quantity
- 1,839
- Product code
- MQP
- FDA submissions
- K112429
- Distribution
- AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WY
Product description
GIZA Vertebral Body Replacement
Code information (models, lots, serial numbers)
Part # 48291482S Cage Giza Diam 14mm, 18-22 mm, 2-5 Deg GTIN 07640167930012 Part # 48291485S Cage Giza Diam 14mm, 18-22 mm, 5-8Deg GTIN 07640167930029 Part # 48291422S Cage Giza Diam 14mm, 23-32 mm, 2-5 Deg GTIN 07640167930036 Part # 48291427S Cage Giza Diam 14mm, 23-32 mm, 7-10 Deg GTIN 07640167930043 Part # 48291432S Cage Giza Diam 14mm, 33-54 mm, 2-5 Deg GTIN 07640167930050 and Part # 48291437S Cage Giza Diam 14mm, 33-54 mm, 7-10 Deg GTIN 07640167930067
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 5-8 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 33-54 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 33-54 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page