Skip to content

Refinity Rotational Intravascular Ultrasound (IVUS) Catheter recallZ-1775-2017 · Volcano Corporation · initiated Mar 29, 2017

Class II Terminated All recalls posted in April 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Volcano Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in disruption of normal clinical workflow or loss of imaging.

FDA root cause category: Under Investigation by firm

What the company is doing

Recall letters dated March 28, 2017 will be hand delivered by Volcano reps. Reps will also pick up any unused catheters for return and obtain a signed proof of delivery for both customer and Volcano Rep.

Recall details

Recall number
Z-1775-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 29, 2017
Posted by the FDA
Apr 10, 2017
Recalling firm
Volcano Corporation
Quantity
718 total units
Product code
OBJ
FDA submissions
K160583
Distribution
Distribution to US and Japan

Product description

Refinity Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.289; Catalog number 89800. Cardiovascular: The REFINITY (or REFINITY ST) Rotational IVUS Catheter (REFINITY or REFINITY ST) is intended for the intravascular ultrasound examination of coronary arteries. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. REFINITY (or REFINITY ST) catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

Code information (models, lots, serial numbers)

Lot # 50068951, Expiry date 06/30/2018; Lot # 50068952, Expiry date 06/30/2018; Lot # 50069133, Expiry date 06/30/2018; Lot # 50083387, Expiry Date 11/30/2018; Lot # 50083390, Expiry Date 11/30/2018.

Other devices under the same FDA submission

1 device cleared under K160583 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 76885).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page