Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part… recallZ-1923-2017 · Merge Healthcare, Inc. · initiated Mar 30, 2015

Class II Terminated All recalls posted in May 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Merge Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

The recall was initiated by telephone on March 30, 2015. Follow-up phone calls were made July 6, 2015. The firm was unable to provide a telephone script. The recalling firm also issued letters dated March 28, 2016, via email on April 6, 2016, dated July 11, 2016. The letters contained the same content, notified the customer of the issue, and informed them a fix was available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they have the affected ISA CO2 sensors at their facility and are they interested in accepting the fix. If they decline the fix, they are asked to provide the reason why. For further questions, please call (877) 741-5369.

Recall details

Recall number
Z-1923-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 30, 2015
Posted by the FDA
May 2, 2017
Recalling firm
Merge Healthcare, Inc.
Quantity
23 sensors (PhaseIn Units)
Product code
DQK
FDA submissions
K082421
Distribution
US Distribution was made to medical facilities located in AL, FL, ID, LA, MO, NC, OK, SC, TX, and UT. There was government distribution and no foreign/military distribution.

Product description

Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.

Code information (models, lots, serial numbers)

ISA CO2 Sensor serial numbers:  107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435,  107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776,  107831, 107832, 107836, 107839, and 107841.

Other devices under the same FDA submission

1 device cleared under K082421 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.