ETEST ETP32 (Ertapenem) Ref. 531640, 531600, blister… recallZ-2095-2017 · BioMerieux SA · initiated Apr 25, 2017
Class II Terminated All recalls posted in May 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioMerieux SA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
False susceptible results
FDA root cause category: Under Investigation by firm
What the company is doing
bioMerieux sent an Urgent Product Safety Correction Notice on April 25, 2017 to U. S. consignees. The letter requested that they provide the notice to all appropriate personnel in the laboratory, asking them to identify any possible False Susceptible results. The letter included a Acknowledgment Form to be returned. On this notice the consignee will indicate the number of kits discarded. For customers with questions please contact their local bioMerieux Customer Service representative.
Recall details
- Recall number
- Z-2095-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 25, 2017
- Posted by the FDA
- May 3, 2017
- Recalling firm
- BioMerieux SA
- Quantity
- 3,238 units
- Product code
- JWY
- FDA submissions
- K020748
- Distribution
- NATIONWIDE DISTRIBUTION TO AL, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, IL, IN, IA, KS, KY, LA, M, MA, MD, MI, MN, MS, MO, MT, NE, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY
Product description
ETEST ETP32 (Ertapenem) Ref. 531640, 531600, blister packaging ETEST is a quantitative technique for determining the antimicrobial susceptibility of Gram-negative and Gram-positive aerobic bacteria (Only 531640 was distributed in the United States)
Code information (models, lots, serial numbers)
REF 531640 Lot Number - 1003149850, 1003157690, 1003585350, 1003732380, 1003910410, 1004142070, 1004113370, 1004325260, 1004382730, 1004482780, 1004785720, 1004890160, 1005134380, 1005134390, 1005357180, 1005442470 --- REF 531600 - was not distributed in the United States
Other devices under the same FDA submission
1 device cleared under K020748 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ETEST® Ertapenem (ETP)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page