Skip to content

ARIES System and ARIES M1 System recallZ-2094-2017 · Luminex Corporation · initiated Apr 19, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Luminex Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The hand-held barcode scanner that is used as an accessory to enter ARIES cassette and specimen identifications on the ARIES Systems populates patient sample IDs with a previously scanned sample ID when entering orders into an ARIES instrument.

FDA root cause category: Nonconforming Material/Component

What the company is doing

A recall notification letter was sent out on April 19, 2017. Prior to the recall letter being sent, the recalling firm sent a Customer Advisory Notification on April 14, 2017. The firm is also adding an accountability form to the customer notifications in order to document the customer responses to ensure that we account for the estimated 15 distributed units. For further questions, please call (877) 785-2323.

Recall details

Recall number
Z-2094-2017
Status
Terminated
Initiated by the company
Apr 19, 2017
Recalling firm
Luminex Corporation
Quantity
15 units
Product code
OOI
FDA submissions
K151917, K160517, K162772
Distribution
Worldwide Distribution - US Distribution and to the countries of Finland and China.

Product description

ARIES System and ARIES M1 System

Code information (models, lots, serial numbers)

ARIES System UDI: 00840487101537 and ARIES M1 System UDI: 00840487100080

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.