Zenith Alpha Thoracic Endovascular Graft recallZ-2099-2017 · Cook Medical Incorporated · initiated Mar 22, 2017

Class I Terminated All recalls posted in August 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cook Medical Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There have been five reports of graft thrombosis/occlusion during global commercial use, each following treatment for BTAI. One case resulted in patient death, and three cases resulted in reintervention.

FDA root cause category: Device Design

What the company is doing

The firm, Cook Medical, sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated March 22, 2017 to the affected consignees via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. On 6/22/2017, updated "URGENT: MEDICAL DEVICE CORRECTION AND REMOVAL" notifications were sent to the affected consignees via courier. The update included removal of the indication for use in BTAI, as follows: The Zenith Alpha Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with aneurysms or ulcers of the descending thoracic aorta having vascular morphology suitable for endovascular repair including: " Iliac/femoral anatomy that is suitable for access with the required introduction systems " Nonaneurysmal aortic segments (fixation sites) proximal and distal to the aneurysm or ulcer: o with a length of at least 20 mm, and o with a diameter measured outer-wall-to-outer-wall of no greater than 42 mm and no less than 20 mm. Additionally, the following warning has been added to describe the thrombus risk that has been observed when the device is used to treat BTAI: " Risk of in-graft thrombus has been observed when the Zenith Alpha Thoracic Endovascular Graft has been used to treat BTAI. Because of the IFU correction to remove BTAI from the indication, it is necessary to remove specific sizes of this device (grafts with a proximal or distal diameter 18 22 mm) that would likely be used only for BTAI. This notice must be shared with appropriate personnel, including down to the user level, within your organization or to any organization where the potentially affected devices have been transferred. Should you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.

Recall details

Recall number
Z-2099-2017
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Mar 22, 2017
Posted by the FDA
Aug 10, 2017
Recalling firm
Cook Medical Incorporated
Quantity
4916
Product code
MIH
FDA submissions
P140016
Distribution
Nationwide Distribution.

Product description

Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descending thoracic aorta (not including dissections) having vascular anatomy suitable for endovascular repair including: " Iliac/femoral anatomy that is suitable for access with the required introduction systems, " Nonaneurysmal aortic segments (fixation sites) proximal and distal to the thoracic lesion: " with a length of at least 20 mm, and " with a diameter measured outer wall to outer wall of no greater than 42 mm and no less than 15 mm.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

All lots. Catalog Numbers: ZTA-D-28-160-W ZTA-D-28-229-W ZTA-D-30-160-W ZTA-D-30-229-W ZTA-D-32-160-W ZTA-D-32-229-W ZTA-D-34-142-W ZTA-D-34-190-W ZTA-D-36-142-W ZTA-D-36-190-W ZTA-D-38-147-W ZTA-D-38-197-W ZTA-D-40-147-W ZTA-D-40-197-W ZTA-D-42-152-W ZTA-D-42-204-W ZTA-D-44-157-W ZTA-D-44-211-W ZTA-D-46-157-W ZTA-D-46-211-W ZTA-DE-18-104-W ZTA-DE-18-148-W ZTA-DE-20-104-W ZTA-DE-20-148-W ZTA-DE-22-104-W ZTA-DE-22-148-W ZTA-DE-24-104-W ZTA-DE-24-148-W ZTA-DE-26-104-W ZTA-DE-26-148-W ZTA-DE-28-108-W ZTA-DE-28-154-W ZTA-DE-30-108-W ZTA-DE-30-154-W ZTA-DE-32-108-W ZTA-DE-32-154-W ZTA-DE-34-112-W ZTA-DE-34-160-W ZTA-DE-36-112-W ZTA-DE-36-160-W ZTA-DE-38-141-W ZTA-DE-38-91-W ZTA-DE-40-141-W ZTA-DE-40-91-W ZTA-DE-42-146-W ZTA-DE-42-94-W ZTA-DE-44-151-W ZTA-DE-44-97-W ZTA-DE-46-151-W ZTA-DE-46-97-W ZTA-P-18-105-W ZTA-P-18-127-W ZTA-P-20-105-W ZTA-P-20-127-W ZTA-P-22-105-W ZTA-P-22-127-W ZTA-P-24-105-W ZTA-P-24-127-W ZTA-P-26-105-W ZTA-P-26-149-W ZTA-P-28-109-W ZTA-P-28-132-W ZTA-P-28-155-W ZTA-P-28-201-W ZTA-P-30-109-W ZTA-P-30-132-W ZTA-P-30-155-W ZTA-P-30-201-W ZTA-P-32-109-W ZTA-P-32-132-W ZTA-P-32-155-W ZTA-P-32-201-W ZTA-P-34-113-W ZTA-P-34-137-W ZTA-P-34-161-W ZTA-P-34-209-W ZTA-P-36-113-W ZTA-P-36-137-W ZTA-P-36-161-W ZTA-P-36-209-W ZTA-P-38-117-W ZTA-P-38-142-W ZTA-P-38-167-W ZTA-P-38-217-W ZTA-P-40-117-W ZTA-P-40-142-W ZTA-P-40-167-W ZTA-P-40-217-W ZTA-P-42-121-W ZTA-P-42-147-W ZTA-P-42-173-W ZTA-P-42-225-W ZTA-P-44-125-W ZTA-P-44-152-W ZTA-P-44-179-W ZTA-P-44-233-W ZTA-P-46-125-W ZTA-P-46-152-W ZTA-P-46-179-W ZTA-P-46-233-W ZTA-PT-22-18-105-W ZTA-PT-26-22-105-W ZTA-PT-30-26-108-W ZTA-PT-32-28-178-W ZTA-PT-32-28-201-W ZTA-PT-34-30-161-W ZTA-PT-34-30-209-W ZTA-PT-36-32-161-W ZTA-PT-36-32-209-W ZTA-PT-38-34-167-W ZTA-PT-38-34-217-W ZTA-PT-40-36-167-W ZTA-PT-40-36-217-W ZTA-PT-42-38-173-W ZTA-PT-42-38-225-W ZTA-PT-44-40-179-W ZTA-PT-44-40-233-W ZTA-PT-46-42-179-W ZTA-PT-46-42-233-W

Other devices under the same FDA submission

67 devices cleared under P140016 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.