Orthofix Power Drill Torque Limiter, Sterile, Rx Only… recallZ-0291-2018 · Orthofix Srl · initiated Jul 5, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Orthofix Srl or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The power drill torque limiter may break during use in surgery application.
FDA root cause category: Process change control
What the company is doing
Consignees were sent an Urgent Medical Device Recall Notification, dated June 28, 2017, informing them of the device correction. The letter instructed distributors to immediately identify affected product in warehouses, quarantine affected product, and communicate the amount on hand to Orthofix SRL for replacement. Any customers are to be forwarded the recall notice provided to the distributor. Distributors are asked to complete the Reply Form with the appropriate information and to return it to Orthofix by fax at +39-045-6719380 or by e-mail to [email protected]. Hospitals were instructed to identify affected product on-site, quarantine affected units, and communicate the quantity of affected product on-hand to their local distributor or to Orthofix. Affected product will be replaced by Orthofix.
Recall details
- Recall number
- Z-0291-2018
- Status
- Terminated
- Initiated by the company
- Jul 5, 2017
- Recalling firm
- Orthofix Srl
- Quantity
- 416 units
- Product code
- KTT
- FDA submissions
- K142052, K153233
- Distribution
- Worldwide Distribution -- US, including the states of CA, FL, & PA; and, the countries of Australia, Germany, Finland, France, Italy, Lithuania, Netherlands, Singapore, South Africa, Sweden, Switzerland, & United Kingdom.
Product description
Orthofix Power Drill Torque Limiter, Sterile, Rx Only, Single Use; Sold individually and in the Galaxy UNYCO Diaphyseal Tibia Sterile Kit, Galaxy UNYCO Mini Kit Instruments Sterile, Galaxy UNYCO Diaphyseal Femur Box, Galaxy UNYCO Knee Bridging Box, Galaxy UNYCO Ankle Bridging Box - Unilateral Frame, Galaxy UNYCO Ankle Bridging Box - Delta Frame, and Galaxy UNYCO Diaphyseal Tibia Box. Intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Power Drill Torque Limiter: Product Code 99-93568; Lot (Exp Date) V1388296 (12/28/2019), V1388432 (12/29/2019), V1389857 (01/13/2020), V1400707 (05/18/2020), V1400940 (05/23/2020), V1410027 (09/01/2020); UDI: 18054242512318. Galaxy UNYCO Diaphyseal Tibia Sterile Kit: Product Code 99-93506; Lot (Exp Date) V1389818 (01/12/2020), V1389819 (01/12/2020), V1389886 (02/13/2020), V1399881 (05/10/2020), V1400800 (05/22/2020), V1402023 (06/03/2020), V1405602 (07/17/2020), V1406422 (07/26/2020), V1410082 (09/02/2020), R1412578 (09/28/2020), V1414451 (10/19/2020), V1421785 (02/01/2021), V1425406 (03/13/2021), B1012462 (06/19/2021), B1012776 (06/26/2021), B1019431 (07/05/2021), B1031608 (08/05/2021), B1052165 (10/23/2021), B1049195 (10/25/2021), B1080028 (02/16/2022), B1099157 (04/08/2022), B1107271 (04/09/2022); UDI: 18054242512325. Galaxy UNYCO Mini Kit Instruments Sterile: Product Code 99-93509; Lot (Exp Date) V1408711 (08/01/2020), V1410026 (09/01/2020), V1422854 (02/14/2021), V1427350 (04/09/2021), V1429217 (04/26/2021), B1083360 (02/20/2022); UDI: 18054242513629. Galaxy UNYCO Diaphyseal Femur Box: Product Code 99-93790; Lot (Exp Date) B1019281 (04/09/2021); UDI: 1805424515159. Galaxy UNYCO Knee Bridging Box: Product Code 99-93791; Lot (Exp Date) B1019282 (04/09/2021); UDI: 18054242515166. Galaxy UNYCO Ankle Bridging Box - Unilateral Frame; Product Code 99-93792; Lot (Exp Date) V140597 (03/21/2020); UDI: 18054242513643. Galaxy UNYCO Ankle Bridging Box - Delta Frame: Product Code 99-93793; Lot (Exp Date) V1409944 (03/09/2020); UDI: 18054242513650. Galaxy UNYCO Diaphyseal Tibia Box: Product Code 99-93794; Lot (Exp Date) V1410194 (04/18/2020). UDI: 18054242513667.
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- ORTHOFIX GALAXY UNYCO DIAPHYSEAL TIBIA KIT 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component1 recall2 reports
- Galaxy UNYCO System 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component1 recall2 reports
- ORTHOFIX GALAXY UNYCO DIAPHYSEAL TIBIA KIT 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component1 recall0 reports
- Galaxy Unyco 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component1 recall0 reports
- Galaxy Unyco 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component1 recall0 reports
- Galaxy Unyco 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component1 recall0 reports
- Galaxy Unyco 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page