Skip to content

Firebird NXG Spinal Fixation System Decorticating Planer… recallZ-3127-2017 · Orthofix, Inc · initiated Aug 2, 2017

Class II Terminated All recalls posted in August 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Orthofix, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or beyond a 23 degree angle.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Orthofix sent an Urgent Medical Device Product Recall letter dated August 2, 2017, to all affected customers. The letter explained the issue and requested return of product in their possession. Customers with questions were instructed to contact their local Orthofix representative. For questions regarding this recall call 214-937-2100.

Recall details

Recall number
Z-3127-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Aug 2, 2017
Posted by the FDA
Aug 18, 2017
Recalling firm
Orthofix, Inc
Quantity
6 planers
Product code
NKB
FDA submissions
K171082
Distribution
Distribution was made to distributors located in CA, FL, and RI. There was no foreign/military/government distribution.

Product description

Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating Planer is used to decorticate bone surrounding the spherical head of an implanted bone screw in order to provide clearance to allow for proper seating of the modular body, as well as to provide full range of modular body angulation. It is placed axially over the spherical head of the bone screw and rotated clockwise and counterclockwise to achieve the needed decortication

Code information (models, lots, serial numbers)

Lot #203013-TD26

Other devices under the same FDA submission

1 device cleared under K171082 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page