Skip to content

Unifine Pentips recallZ-0381-2018 · Owen Mumford USA, Inc. · initiated Nov 6, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Owen Mumford USA, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device may potentially be compromised by water in end user packaging. Exposure to liquid may compromise sterility. This may appear as discoloring or yellowing of the device's protective seal.

FDA root cause category: Environmental control

What the company is doing

Owen Mumford sent an Urgent Medical Device Recall Notification letter dated November 6, 2017 via email. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to check inventory for affected product, contact the Customer Service Department to arrange return of affected product, complete and return the response form and to either provide a customer list for Owen and Mumford to notify their customers or confirmation that they will notify their customers if further distributed.

Recall details

Recall number
Z-0381-2018
Status
Terminated
Initiated by the company
Nov 6, 2017
Recalling firm
Owen Mumford USA, Inc.
Quantity
2,660 cartons (266,000 pen needles)
Product code
FMI
FDA submissions
K152339
Distribution
US Nationwide Distribution in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, MA, MI, MN, MO, MS, NC, NH NJ, NY, OH, OK, PA, TN, NX, UT, VA, VT, WA, WV

Product description

Unifine Pentips 5mm x 31G pen needles, packaged in 100 ct pen needles. Product Usage: Sterile, non-toxic, non-pyrogenic, single use pen needles for the subcutaneous injection of drugs from pen injectors.

Code information (models, lots, serial numbers)

UPC #384703550018: Batch No. AN 3550, Lot No. 1705722.S846; Batch No. AN 3550AC, Lot No. 1705122.S929.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

17 devices cleared under K152339 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 78483).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page