BRAINLAB EXACTRAC VERO, Model/Catalog Numbers recallZ-0172-2018 · Brainlab AG · initiated Nov 3, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Brainlab AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This may result in the treatment couch not moving to the exact planned position resulting in a deviation between the planned and the treatment target position. This deviation may not be clearly visible to the user if no verification image of the patient position is acquired after this couch shift has been performed.

FDA root cause category: Software design

What the company is doing

The firm notified their consignees by email on November 3, 2017. "Instructions to customers what to do with the recalled product. For instructions for the user how to address this issue, and to avoid a patient and treatment target position potentially different than intended due to this issue, please also refer to the User Corrective Action section in the attached Field Safety Notice / Product Notification letter CAPA-20171011-001985: 1. After applying a couch shift always perform a verification image of the patient position before enabling the treatment beam. 2. Do not use the Virtual Isocenter function if a verification of the patient position in the treatment isocenter is not feasible. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning in the Clinical User Guide ExacTrac Vero: Brainlab highly recommends performing an X-ray verification after each change to the couch/Robotics Position." For further questions, please call 1 (800) 784-7700.

Recall details

Recall number
Z-0172-2018
Status
Terminated
Initiated by the company
Nov 3, 2017
Recalling firm
Brainlab AG
Quantity
28 units
Product code
IYE
FDA submissions
K122451
Distribution
Worldwide Distribution - US Distribution and to the countries of : Belgium, Canada, France, Germany, Italy, Japan and S. Korea

Product description

BRAINLAB EXACTRAC VERO, Model/Catalog Numbers: 46228 EXACTRAC VERO 3.5 46238 EXACTRAC VERO 3.5 46216 EXACTRAC VERO SW UPDATE 3.5.2 TO 3.5.3 46218 EXACTRAC VERO SW UPDATE 3.5.3 TO 3.5.4 The ExacTrac Vero system is intended to be used in conjunction with the MHI-TM2000 Radiation Therapy Linear Accelerator System manufactured by Mitsubishi Heavy Industries, Ltd. ExacTrac Vero uses the images received from the MHI-TM2000 linear accelerator for analyzing the current patient position and calculating - when applicable - a necessary correction shift. The correction shift is then exported to the MHI-TM2000 linear accelerator. The ExacTrac Vero system uses stereoscopic X-ray or Cone Beam CT registration and optical tracking of infrared reflective markers in order to localize and correct the patient position before and during treatment. Optionally ExacTrac Vero provides position data for the pan/tilt motion of the MHI-TM2000 gantry head to the MHITM2000 controller for continuous alignment of the beam orientation with a moving target. The position data is based on target detection via X-ray imaging and IR tracking of external surrogate markers.

Code information (models, lots, serial numbers)

Potentially affected are ExacTrac Vero versions 3.x. Currently only ExacTrac Vero SW versions 3.5(.x) are installed for use on existing ExacTrac Vero Systems (only ExacTrac Vero v.3.5 exist in the field). UDI: 04056481132446

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.