Why it was recalled

Particulate found on the Curlin spike.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The recalling firm began issuing letters via USPS certified mail with return receipt on 11/10/2017 and concluded on 11/13/2017. Updated letters dated 11/20/2017 were issued via certified mail to distributors and to medical facilities. The distributor letter emphasized their instructions to notify downstream accounts. The medical facility letter updated lot numbers. The letter described the issue, risk to health, and actions the customer was to take. A response form was enclosed to report whether the customer has inventory they are not returning, that they no longer have inventory of the product, or the amount of product being returned for replacement. The form was to be returned via FAX or email.

Recall details

Recall number
Z-0697-2018
Status
Terminated
Initiated by the company
Nov 10, 2017
Recalling firm
Zevex Incorporated (dba Moog Medical Medical Devices Group)
Quantity
56,244/20 administration set cases
Product code
FPA
FDA submissions
K981816
Distribution
Distribution was nationwide. There was government and military distribution. Foreign distribution was made to Canada, Saudi Arabia, Turkey, Croatia, United Arab Emirates, and Australia.

Product description

MOOG Curlin Infusion Administration Set, Ref 340-4114, Non-DEHP Microbore Tubing with Non-Vented Bag Spike, packaged 20/box, for use with the MOOG Curlin Ambulatory Volumetric Infusion Pump, RX, Sterile. Product Usage: Administration set to be used with the Moog Curlin infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral infusions.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI 38148440000396, Lot/serial numbers: CF1514703, CF1514702, CF1514702, CF1514701, CF1514702, CF1517405, CF1517406, CF1518301,  CF1518302, CF1518303, CF1518301, CF1521101, CF1521103, CF1521102,  CF1521701, CF1522306, CF1522308, CF1522307, CF1523312, CF1524407,  CF1524407, CF1525101, CF1524407, CF1525103, CF1525102, CF1526505,  CF1526506, CF1527307, CF1527306, CF1528011, CF1528607, CF1528606,  CF1529306, CF1529307, CF1530106, CF1530707, CF1530706, CF1530706,  CF1530705, CF1530706, CF1530707, CF1530706, CF1530707, CF1532102,  CF1531406, CF1532102, CF1531405, CF1532975, CF1532976, CF1532976,  CF1532976, CF1532976, CF1535001, CF1533504, CF1534902, CF1535666,  CF1600507, CF1600508, CF1601203, CF1601903, CF1602604, CF1603304,  CF1604704, CF1605406, CF1606104, CF1607504, CF1610405, CF1611003,  CF1612404, CF1613105, CF1613804, CF1615209, CF1616608, and CF1616608.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 78573).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page