PENTARAY NAV recallZ-0212-2018 · Biosense Webster, Inc. · initiated May 17, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biosense Webster, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Contraindication language updated in the Instructions For Use (IFU) to indicate that Pentaray Catheters should not be used in patients with prosthetic valves.
FDA root cause category: Other
What the company is doing
Recall Notification letter, dated 04/14/2016, and Acknowledgement Form was sent to customers via express mail on 05/17/2017. Instructions included to read and follow Labeling Correction Notice carefully, provide the notice to anyone in the facility that should be aware, sign and return the Acknowledgement Form, and to maintain awareness of the correction.
Recall details
- Recall number
- Z-0212-2018
- Status
- Terminated
- Initiated by the company
- May 17, 2017
- Recalling firm
- Biosense Webster, Inc.
- Quantity
- 73,796 units total
- Product code
- MTD
- FDA submissions
- K120425, K123837
- Distribution
- Nationwide Distributed internationally to Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Republic of Korea, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, The Netherlands, New Zealand, Norway, Oman, People's Republic of China, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
Product description
PENTARAY NAV High-Density Mapping Catheter.
Code information (models, lots, serial numbers)
Product code (Product Description) (GTIN): D-1282-07-S (NAV ECO 7FR, F, 4-4-4) (10846835012217), D-1282-08-S (NAV ECO 7FR, F, 2-6-2) (10846835012224), D-1282-09-S (NAV ECO 7FR, F, 1-8-1) (10846835012231), D-1282-10-S (NAV ECO 7FR, D, 4-4-4) (10846835012248), D-1282-11-S (NAV ECO 7FR, D, 2-6-2) (10846835012255), D-1282-12-S (NAV ECO 7FR, D, 1-8-1) (1846835012262).
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- PENTARAY NAV D128211Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls6 deaths551 reports
- PENTARAY NAV D128208Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls2 deaths237 reports
- PENTARAY NAV D128210Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls31 reports
- PENTARAY NAV D128207Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls15 reports
- PENTARAY NAV D128209Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls0 reports
Product lines
Other devices under the same FDA submission
7 devices cleared under K120425, K123837 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- PENTARAY NAV D128201Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls6 reports
- PENTARAY NAV D128203Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls1 report
- PENTARAY NAV D128205Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls1 death1 report
- PENTARAY NAV D128202Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls0 reports
- PENTARAY NAV D128204Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls0 reports
- PENTARAY NAV D128206Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls0 reports
- PENTARAY NAV D128212Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 77529).
- Z-0211-2018 · TerminatedPENTARAY NAV High-Density Mapping Catheter.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.