Skip to content

Brainlab Navigation System Spine & Trauma 3D with component… recallZ-0320-2018 · Brainlab AG · initiated Nov 16, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Brainlab AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a human error at a supplier, an unknown number of devices within one specific production lot of Brainlab DrapeLink Adapters (Left/Right) were not manufactured according to defined specifications.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Customers were notified, via email on approximately 11/16/2017. Instructions included to identify all affected devices on hand, cease use of the affected device, and a Brainlab Customer Support Representative will contact the customer to arrange a service visit to replace the item

Recall details

Recall number
Z-0320-2018
Status
Terminated
Initiated by the company
Nov 16, 2017
Recalling firm
Brainlab AG
Quantity
80 devices total
Product code
HAW
FDA submissions
K070106
Distribution
Distributed in the US to Kansas. Distributed internationally to Australia, Finland, France, Germany, Hong Kong, Ireland, Japan, Netherlands, Norway, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.

Product description

Brainlab Navigation System Spine & Trauma 3D with component Brainlab DrapeLink Registration Kit for C-arm Zeihm Vision RFD: Adapter Left

Code information (models, lots, serial numbers)

Model No. 19154, UDI GTIN 04056481132187. Model No. 19154-04, UDI No. GTIN 04056481113926. Serial No. range 7459316001 - 7459316027 and 7714816001 - 7714816017.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K070106 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates