Brand Name recallZ-0378-2018 · Roche Diagnostics Corporation · initiated Jun 29, 2017
Class II Terminated All recalls posted in January 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Diagnostics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Roche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.
FDA root cause category: Software design
What the company is doing
Roche sent an Urgent Medical Device Correction (UMDC) sent via UPS Ground (receipt required) dated July 6, 2017, and an amended UMDC on August 31 ,2017. All consignees will be notified by Urgent Medical Device Correction sent via UPS Ground (receipt required). Customers/consignees were instructed the following: If the Tip/Cup pick up error alarm is issued then followed by the Abnormal L2-line Movement alarm on your MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer, after which the system goes into Sampling Stop, follow the steps in this UMDC and the appropriate How to Identify Potentially Mismatched Sample Results within the enclosed document. For further questions, please call (317) 521-4343.
Recall details
- Recall number
- Z-0378-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 29, 2017
- Posted by the FDA
- Jan 16, 2018
- Recalling firm
- Roche Diagnostics Corporation
- Quantity
- 2517
- Product code
- JJE
- FDA submissions
- K060373, K100853, K953239
- Distribution
- USA (nationwide) Distribution
Product description
Brand Name: cobas c 6000 MODULAR Series System e 601 cobas c 8000 MODULAR Series System - e 602 Roche / Hitachi MODULAR Analytics Combination System Model/Catalog/Part Number: MODULAR Analytics E 170 03022109001 03023109973 03739040001 03739040692 05023572001 05023599001 cobas e 601 module 04745922001 04745922692 05036348001 05036348692 05860652001 cobas e 602 module 05990378001 The Roche/Hitachi MODULAR Analytics System (E170) is a fully automated, random-access, software- controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. The cobas 6000 series (e 601) is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. It is optimized for high throughput workloads in the professional environment using a combination of Ion selective electrodes (ISE), a photometric analysis unit an immunoassay analysis module. The cobas 8000 Modular Analyze Series (e602) is a fully automated system for Clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit.
Code information (models, lots, serial numbers)
Show the full code information (20k characters)
Serial Number 2072-26 27P6-20 15S2-20 2717-18 2719-02 1544-11 1686-03 1534-12 1698-04 1549-04 2275-12 2263-16 2263-17 2497-16 1849-01 1570-15 26F0-17 2256-18 2564-30 2850-05 2384-12 2384-18 2399-08 2230-19 2230-20 2379-14 2381-02 27Z8-01 2058-22 1909-13 1909-01 27V5-20 15U5-18 2308-01 2087-07 1984-05 1985-07 26D3-30 2497-01 1861-18 1861-19 26D0-27 26D1-21 1480-14 1480-15 1488-01 1488-02 26C3-12 26F0-19 26F0-27 26D9-24 2107-35 2348-12 2348-13 1423-16 2237-18 2240-09 2240-10 16J8-05 16J8-06 16L7-20 16K2-17 16K2-18 16K2-19 16K2-20 16K3-01 16K3-06 16K3-07 16K3-08 16K3-09 16L4-03 16L4-04 26D1-25 26D1-26 26D1-28 26D1-29 2328-03 2437-13 2072-29 2072-30 2064-12 1931-11 2080-35 1936-08 2049-15 27Z0-09 27Z0-10 1911-11 2581-18 2062-14 2074-05 1430-16 1430-17 1691-17 1691-18 1985-13 1989-19 1739-19 2578-25 13B8-12 1169-06 1169-08 1169-09 1211-16 1211-17 1211-18 1224-10 1458-04 2543-10 27J6-15 2048-05 2406-27 26B3-30 1927-19 1914-13 1914-11 27J7-06 2822-07 2822-08 2535-23 2306-04 1159-19 2867-11 2711-02 27Y6-19 2107-20 2195-34 2106-20 2079-34 27M5-13 15D1-02 15C9-10 15F0-17 15F0-18 2137-15 1872-01 2195-40 2102-30 2428-02 2539-08 1377-14 2495-01 1490-16 15C5-16 15C5-17 15C5-20 15C6-01 2250-02 26H6-29 2423-15 2539-06 1377-12 2307-26 2307-27 2227-11 2080-24 1128-05 1128-06 14B1-02 14B2-13 2467-05 27Z8-09 26E4-05 2413-14 2423-19 2086-27 27Z0-14 27Z0-16 2145-16 2321-26 2321-25 26F1-20 26F1-19 1305-40 26G7-23 26G8-09 27L8-12 27X4-14 2719-07 27S8-20 15U1-20 26A4-28 26D1-22 2368-06 1128-19 2427-29 2539-07 1377-13 1357-08 1357-09 1357-10 1344-05 1344-06 14B1-01 1344-02 1344-03 1344-04 1469-03 1469-04 1469-05 1476-20 1482-11 1483-01 1483-17 1483-18 1483-19 1483-20 1484-16 1484-17 1484-18 1490-18 1490-19 1490-20 14A3-01 14A3-02 14A3-03 15B7-01 15B7-02 2430-27 2430-29 15B7-03 2327-25 2347-04 2369-04 2332-04 2332-09 27S4-02 27S4-03 2230-14 2230-18 1211-15 1211-14 15D2-04 15D2-05 15D3-12 15F0-15 15F0-16 16P4-02 16P4-03 2061-21 14A9-16 14A9-17 14A9-18 15B7-04 2256-19 2865-20 2867-01 2070-11 2070-13 2720-17 16X4-17 2102-29 2192-37 2196-02 2535-18 2842-03 2896-04 16P4-04 16P4-07 16P4-08 16P4-09 2311-12 2311-13 1871-09 1125-07 1128-02 2059-28 15H8-11 15H8-12 1996-01 2396-04 2399-06 2068-24 2283-14 2080-02 2577-04 2047-10 2107-05 26E4-10 27J7-04 2585-07 2581-08 2581-13 2327-11 2396-07 2413-01 2413-04 185731 1867-06 2037-07 2423-14 1549-02 27Z8-02 1860-19 2074-25 2087-09 2103-35 1924-15 2192-02 2437-19 2437-20 27Z0-03 2437-17 1913-03 1913-04 2488-05 2402-05 26C3-11 26C9-12 26E4-08 26F3-21 1545-05 1844-14 2822-09 2822-10 1303-31 1303-32 1357-19 1357-20 2402-30 2195-38 2196-01 2080-11 2080-17 2226-17 2439-12 2107-02 1988-03 2086-40 2195-24 2195-06 2195-08 2072-39 2041-12 2047-07 2192-15 2192-08 1736-04 1991-08 1871-06 2037-14 2055-30 2055-35 2873-02 2891-13 1990-05 192506 2894-12 1936-14 1928-19 2848-17 2867-08 2058-21 2058-25 2073-01 2111-04 130833 130834 2850-06 1334-12 1334-13 1334-14 1476-01 1476-02 1476-12 2049-29 2049-30 28E5-14 2347-11 2367-09 27Z8-12 2842-02 192418 1936-09 28A2-10 28AO-16 2208-17 2227-10 15J5-02 15J5-03 1852-15 2256-15 1461-03 1461-04 1476-11 16M7-01 16M7-16 14B3-06 14B3-07 16K3-15 16K3-16 16K3-17 16K3-18 16K3-19 16K3-20 16K4-07 16K4-11 16K4-20 16K5-01 16K5-02 16K5-03 16K5-04 16K5-05 16K5-09 16K5-10 16K7-04 16K7-05 16K7-06 16L4-05 16L4-06 16L4-07 16L4-08 16L7-19 16L5-03 16L7-17 16P1-13 16P1-14 16P1-15 16P2-19 1417-20 2581-25 2587-14 1431-14 2065-12 1545-03 16M7-13 16M8-01 16M8-02 16P1-11 2275-13 2267-19 2275-09 2275-10 2056-14 2047-08 185715 185817 26C3-13 2137-09 15D9-17 15D9-18 2486-24 2489-30 2054-06 2054-10 26E0-18 2467-15 16J8-12 16J8-13 1871-07 2567-22 2367-13 2304-14 2304-15 2202-07 2288-10 1487-16 1488-03 1480-16 1488-04 2427-27 1857-13 2486-06 27J7-03 27Z0-17 27Z0-18 27Z0-19 2494-30 2071-34 2576-22 26E4-06 1124-05 1124-06 1132-03 1132-04 16K5-06 2493-18 2497-12 2848-14 2848-15 2842-01 2848-12 2896-03 2263-13 2263-14 2050-29 2547-29 2550-19 2058-17 2058-20 2054-07 2054-08 2328-05 2439-13 1535-09 2548-03 184826 2497-13 2039-09 2061-35 1905-11 0753-22 1685-16 1570-14 2135-17 2140-11 2493-28 2493-29 2850-09 2850-10 2581-07 2581-22 1867-14 27P6-19 15S2-19 26G7-24 26G7-25 26H3-28 2108-36 192717 2493-23 2494-21 1225-14 2822-20 2824-01 2585-04 1910-02 2540-23 2540-24 1694-19 2104-01 2047-36 0884-01 1991-18 2389-18 2107-29 2108-27 26D6-03 26D5-22 1174-08 2189-02 2189-03 2086-13 2086-21 2446-13 1557-12 2486-23 2867-03 191007 2137-14 16K5-12 16K5-13 16K5-14 16K5-15 2104-02 1725-20 26E4-04 26A9-09 2869-20 16Z2-20 2581-26 2389-13 1142-15 2075-40 0817-20 1927-15 1156-01 1160-04 1156-02 1156-09 1160-06 1331-10 1331-11 1331-12 2072-34 1548-08 2262-18 2405-26 16M8-03 16R5-08 15B7-12 15D2-01 15D2-02 15D2-03 16P2-20 2720-20 2451-02 2451-06 2451-09 2468-18 1985-12 1860-15 2850-03 2850-04 2067-31 2086-28 0829-17 191305 2107-11 1848-01 1848-03 27Z0-15 2704-19 2047-39 0884-04 1421-11 2423-16 2423-17 1482-12 28E7-02 2413-02 2413-03 2486-05 2074-17 2486-04 2486-16 2327-23 2398-05 2438-15 2438-16 2479-09 2479-11 2311-07 2311-11 2425-11 2427-18 2227-07 2233-16 27X4-18 2427-21 2381-03 2382-04 2822-19 2867-07 2850-01 2382-06 2227-02 26C7-30 14H8-04 1017-06 1021-03 1021-05 2108-31 2108-32 2414-14 2367-11 2367-12 2451-19 2406-28 27M5-12 2368-20 2369-01 2398-14 26G4-01 26G4-09 1686-01 2369-05 16S0-01 2080-19 1874-06 26D8-26 26E4-14 26D5-24 26D4-26 26D4-27 2039-07 2585-27 26A4-26 2570-30 13A9-09 2240-13 2240-15 1914-18 2072-31 2072-18 2304-13 2362-19 1122-18 1924-13 2049-14 16R5-15 27L8-11 1868-08 2446-12 2535-20 28E7-16 2548-24 1387-17 1486-20 1490-17 14A3-04 14A3-05 14A3-06 14A3-07 1118-13 2534-05 2534-06 1872-12 2047-37 0884-02 2413-09 2891-15 26H1-28 2858-02 16J8-16 16J8-17 16K2-01 27P2-10 2704-20 2441-10 2441-11 2422-18 2425-12 2219-18 2234-07 1988-11 26B5-23 2379-13 2406-29 2407-04 2894-20 2896-01 1919-05 2102-31 1872-07 1476-03 1479-07 2576-21 185503 27S2-18 2228-15 1005-29 2413-11 2429-30 1263-20 27P2-16 27P2-17 2108-05 2108-21 2425-14 2039-13 2041-16 2413-06 26F1-17 27S4-04 27S4-05 2585-19 2585-21 2493-05 2493-19 1910-04 2422-29 26D6-02 26E0-16 2842-11 2107-31 2848-18 2193-39 27R3-06 27R3-08 27L1-09 2842-10 1914-20 0763-09 2347-15 2348-16 27Z8-08 1691-15 2427-25 2427-26 2207-08 2226-09 26C7-29 14H8-03 2577-02 2577-01 26D4-29 26D4-30 1927-09 2848-16 1459-12 1459-13 1476-05 1476-06 2073-10 26C3-07 190703 2049-21 2369-03 2587-10 2305-22 2070-21 27J7-09 2066-27 2103-40 26H3-26 2848-19 2061-36 0801-27 2581-23 2584-23 2584-30 2395-29 2396-02 1870-16 2073-38 2080-30 2578-24 13B8-11 2071-20 2071-21 2065-04 27W3-18 1344-01 1344-07 1344-08 2288-12 2288-07 27Z8-18 27Z8-19 28E7-13 28E7-14 2140-06 2140-07 2257-02 2079-30 16P1-10 1479-08 1476-07 1476-08 1476-09 1476-10 1988-02 2137-17 14A9-01 14A9-05 2563-30 27R8-19 15S9-19 26B4-26 27J7-02 26H6-26 27Y6-16 27S0-18 27S0-19 26D9-30 2473-07 2486-21 27M5-14 2368-03 1128-16 2319-06 28A3-02 2363-01 26H5-28 28D3-18 1122-20 14M7-18 1616-18 16X4-19 2711-01 2232-19 2708-15 26D6-04 26D8-24 1872-15 1873-05 2054-11 2193-11 2140-03 1855-33 1927-06 1927-20 2074-15 2107-07 2074-14 2209-20 2078-16 2085-33 2137-11 2347-16 1914-17 2358-18 1914-15 0763-06 1118-12 0763-04 15F5-15 15F5-16 2077-38 2078-20 15H2-05 16K4-05 2074-30 2408-22 2580-28 2580-29 1679-13 2497-02 2497-03 26A9-06 1138-10 1139-01 1127-08 1138-08 16M7-05 1988-06 2148-15 28E5-13 2486-27 2493-04 1912-20 0761-08 2374-11 1684-07 1684-08 1570-17 1479-06 1480-17 15F5-12 15F5-13 15H2-08 15H2-09 15D9-15 15D9-16 1691-19 1548-16 2126-08 15H7-15 15H7-16 15H7-17 15H7-18 2068-33 1857-12 2850-15 2106-17 2858-03 2858-04 27P2-07 27P2-08 26D8-20 1418-03 1418-04 1561-12 1925-04 16P4-11 16P4-12 16P4-16 16P4-17 16P4-18 2187-36 2187-37 2187-38 2331-01 2085-09 2086-17 2086-18 2086-15 2086-16 2086-19 2188-08 2188-09 2188-10 2188-36 2188-37 2188-38 2102-06 2102-07 2102-08 2253-16 2253-17 2253-18 2260-11 2260-12 2260-13 26H6-24 2493-27 1858-04 1681-09 1860-21 1861-24 2065-10 1860-20 1860-17 2050-04 2566-26 27P0-20 2850-16 16K6-11 16K6-12 16K6-13 16K6-14 16K6-15 16K6-16 2061-20 2085-12 2085-14 2283-15 2065-06 2081-07 2081-08 2081-09 2081-10 2081-11 2083-04 2083-05 2084-22 2084-23 2083-06 2083-07 2083-08 2188-31 2188-34 2188-35 2192-21 2192-22 2192-26 2192-27 2260-17 2260-18 2260-19 2260-20 2262-11 2262-12 2260-14 2260-15 2260-16 2071-25 2075-20 2263-15 2283-18 2451-17 2451-18 2347-19 2348-11 28E5-12 1733-15 2052-01 2052-02 16K6-17 16K6-18 16K6-19 2086-25 2086-26 2189-07 2082-39 2084-05 2084-12 2084-13 2084-14 2084-15 2084-06 2084-07 2192-31 2101-39 2257-08 2257-09 2257-10 2257-11 27S2-05 27S2-06 2050-06 2050-07 15D6-18 15D9-12 15D9-13 15D9-14 2071-22 2086-23 2086-24 2252-05 2255-20 2263-12 2084-08 2084-09 2084-10 2084-11 2085-04 2085-10 2188-04 2187-33 2188-39 2188-40 2102-03 2195-23 2146-17 2146-18 2146-19 2253-19 2253-20 2254-01 2257-04 2577-03 16K5-17 1616-17 27Z0-07 1905-05 1307-41 2050-01 1544-13 1536-04 2085-11 2084-19 2084-20 2084-21 2413-08 2891-14 2446-14 2446-16 1537-15 1989-20 2050-08 15H7-11 15H7-12 15H7-13 2082-40 2083-01 2084-16 2084-17 2084-18 2188-32 2189-37 2189-38 2189-39 2251-16 2251-17 2251-18 2452-01 2469-06 2232-18 2234-02 26C3-08 26C3-09 2488-02 2232-20 2039-10 1357-15 1357-16 1357-17 1357-18 26G8-08 27R8-20 15S9-20 2493-13 2493-15 2469-05 2469-07 2234-01 2348-17 2145-17 2850-12 2850-13 2145-11 2145-12 15D3-13 15D5-04 15D5-06 15D6-17 1308-35 1308-36 1419-12 1422-17 16J8-14 16J8-15 2079-18 2081-12 2192-18 2147-05 2147-06 2085-05 2085-07 2085-08 2190-14 2188-01 2119-14 28D7-19 1616-19 1921-01 1921-02 1537-17 1905-01 1905-02 2050-03 2050-05 15E3-04 15E3-05 15E3-06 15E3-01 15E3-02 15E3-03 2085-16 2085-17 2189-01 2189-05 2189-06 2256-01 2278-09 2081-06 2188-05 2188-06 2188-07 2190-17 2190-18 2190-21 2190-03 2190-04 2190-05 2189-40 2190-01 2190-02 2189-36 2192-24 2086-12 2106-28 2106-29 2110-09 2110-10 2188-02 2119-13 2056-21 2147-04 2251-15 2253-12 2253-13 2253-14 2257-05 2257-06 2257-07 2059-05 2140-08 2140-10 2473-02 2473-03 2076-22 1544-10 1733-18 1725-19 1857-27 1556-12 1681-08 1694-20 16M6-20 16M7-02 16M7-03 16M8-04 2076-19 2076-20 2076-21 2083-03 2085-18 2083-02 2085-13 2275-14 2275-15 2275-16 2278-07 2190-13 2190-15 2190-16 2192-28 2192-29 2290-05 2290-06 1711-16 2192-12 2189-04 2086-14 2086-20 2190-10 2190-19 2190-20 26E4-12 26H6-23 27L7-20 2069-01 2087-03 26D8-21 26D8-25 2873-01 2891-18 1921-03 1557-10 1736-07 1855-12 1905-03 1925-03 16J8-07 16J8-08 16J8-09 16J8-10 16J8-11 2080-03 2085-15 2252-04 2252-03 2278-08 2079-19 2080-01 2081-04 2190-06 2190-07 2190-08 2190-09 2190-11 2190-12 2192-25 2192-32 2252-02 2253-15 2251-19 2251-20 2283-16 2296-20 27S8-10 27S4-06 2486-14 2486-15 27R3-20 27R5-01 2869-18 16Z2-18 2107-04 2485-29 2487-28 2587-12 2145-08 2290-12 2486-02 2062-02 2467-03 2066-22 2066-23 2102-36 2708-19 2719-08 27R3-05 1556-16 1861-23 2581-21 2066-35 2066-36 26G4-18 26G4-11 26G7-22 26C9-13 26C9-14 2070-15 2472-10 2481-09 2195-36 1160-01 1160-05 1183-04 1217-09 1225-11 26E4-11 2212-19 2212-20 1864-17 1864-18 2427-13 2431-26 26D6-06 14A8-02 16L5-02 16L8-01 16L8-02 16L8-03 16P4-10 27M5-15 2399-09 2399-13 2399-04 2087-08 2585-18 1459-10 1459-11 2104-04 2104-07 2472-11 2472-12 2423-18 27X6-02 27X6-03 27X6-04 27J7-10 27S2-17 27S2-20 2493-26 2212-12 2212-13 2232-13 2232-14 2232-15 26E0-19 2145-13 26B5-24 26C3-05 26C3-06 2107-30 2137-12 2137-13 2581-16 2441-04 1684-05 1572-19 2192-10 2103-29 2104-06 2356-01 27P0-08 27S0-16 1571-13 1867-19 1935-18 1908-03 0756-30 1139-02 1139-03 1139-04 1139-05 2427-19 2427-20 2497-14 2305-23 2307-04 27S2-19 2543-04 27Z0-04 2059-23 2382-05 26G6-30 2358-14 2570-29 1118-08 13A9-08 2348-20 26E4-07 1684-06 2057-28 2487-30 1873-03 2535-22 2535-21 2049-16 27S0-20 186225 2395-28 2865-17 2533-21 1910-08 27S4-01 27P2-01 2485-30 26D8-23 26H6-25 2367-10 2439-14 2486-18 2389-15 1142-17 1860-22 2389-12 1142-14 2865-19 1310-20 1871-12 2069-24 2250-08 2256-14 27S2-01 2227-14 2547-30 2548-01 1925-07 2413-15 2565-21 2565-22 2577-05 2585-22 1556-17 1843-02 26F0-29 27L8-01 27L8-08 2409-26 26A4-27 26B6-30 14G5-06 2565-23 2565-28 2073-27 2079-39 2427-23 2427-24 2544-21 1383-03 2566-25 2566-30 2486-07 2402-28 2226-16 2226-18 26D6-11 26D8-28 2585-25 27M5-11 26A4-23 26A4-29 26A5-01 2587-13 26D0-25 26D0-26 26D5-30 26D9-25 26E0-14 26G4-13 2585-20 2432-29 2432-30 2432-24 1211-19 1211-04 1211-05 1335-08 1335-09 1335-10 1335-11 1910-06 2438-14 27L1-05 2077-40 2078-21 2535-12 2535-13 2535-14 2550-12 2550-14 2565-25 2565-26 27J6-19 2550-15 2564-23 2565-24 2108-30 14B1-03 14B1-04 2423-03 26B4-22 2072-40 2107-38 2229-12 2233-20 2450-17 2451-01 2446-17 2446-19 2406-25 27Z0-12 27P6-16 27P8-01 16R0-06 16R5-06 2581-14 2581-15 2706-01 2493-25 2867-09 2871-19 27R1-16 2355-01 2485-28 1931-09 2039-11 2224-17 2473-04 2473-06 2486-17 2714-09 2714-10 2458-23 1387-16 26F3-20 26G4-12 2581-17 2437-18 27J6-17 2486-13 27Z6-18 27Z6-19 2472-08 2472-09 27S2-15 2720-16 16X4-16 26G4-10 16K3-11 16K3-12 16K3-13 26H3-25 2427-16 2427-17 26H5-29 14M7-19 2396-18 2321-23 2321-24 2362-20 1122-19 1146-01 1146-02 1146-03 16K3-10 2071-40 2219-17 2229-14 2535-06 2535-07 27W3-19 27W3-20 2398-15 2398-16 2407-06 26D4-28 26D8-22 2256-17 26D0-21 26D1-30 2061-29 1139-06 1139-07 15H2-11 15H2-12 15H2-14 2585-06 28E7-12 28F7-02 1132-05 1132-10 1137-02 2423-21 2425-13 26H0-30 14M0-20 2350-04 2408-23 2409-23 2432-25 1924-12 1935-17 2073-37 2042-03 2497-11 2140-12 2704-16 27L1-15 2220-06 2444-07 26F0-18 2080-25 2106-10 28E9-12 28F5-13 16R5-07 14B1-07 15B7-07 15B7-13 2842-14 2842-15 1853-03 1860-16 1733-16 1736-03 2070-01 2070-02 1601-03 1705-20 1208-05 1211-07 1211-08 2330-20 2409-24 2409-25 2107-18 1139-08 1139-10 1140-07 1140-10 1140-08 1156-06 1158-09 1156-07 14B3-08 14B3-09 14B3-10 2369-02 2564-21 2564-22 2227-09 1572-20 1673-01 27Y6-11 27Z0-06 27Z0-11 2413-12 2413-13 1928-01 1848-04 1925-08 1909-09 1909-10 1926-03 2567-23 2423-05 1936-10 1171-10 1211-13 1988-12 1983-06 2494-22 2054-09 1563-04 26B4-21 16K7-12 16M7-06 2856-19 2869-17 2856-20 16Y8-19 16Z2-17 16Y8-20 16L3-20 16K7-11 1866-13 2427-30 2428-01 26A5-08 26A9-03 2304-30 2305-08 2137-06 2564-29 1984-01 2384-19 26A4-24 26A4-25 2048-06 2048-29 2050-28 2072-19 26D6-07 26D6-08 26D6-10 15F0-14 15H2-06 15H2-07 15H5-01 2423-22 2867-05 2867-06 2894-07 2867-12 2894-10 2894-11 16K5-16 16K5-20 16M7-07 16M7-08 2867-13 2585-05 2585-26 2145-14 2145-15 2080-26 27X4-19 27X6-01 2398-17 2396-01 26A9-07 15H4-08 15H4-18 15H5-02 15H5-03 16K5-17 16K5-19 2069-08 2070-05 2313-17 2871-20 16R5-09 16R5-10 16R5-11 27L8-13 1992-19 1999-16 1999-17 2550-10 2550-11 1865-05 1908-10 0757-07 2867-10 16P1-06 16P1-07 16P1-08 16P1-09 2306-02 2306-03 1159-17 1159-18 27P2-05 1927-08 2219-20 16J9-20 1611-20 2312-26 2312-28 2444-08 1987-12 28F5-17 2493-08 2497-17 2422-17 26G4-16 2137-16 2896-02 27P2-04 2079-27 2079-36 2565-27 2567-21 2074-04 2074-01 2538-29 15B7-05 15B7-09 15B7-10 26B3-29 1991-07 1860-23 1871-11 2059-11 27L1-14 27Z8-13 2550-04 2896-05 2060-25 2066-39 26D5-28 16M7-19 16M7-20 1173-01 1985-06 2842-13 27Z8-04 27Z8-05 1919-07 1919-08 2560-03 13A1-02 2256-16 1919-03 1911-15 2050-31 2050-32 1909-08 2074-13 26G4-15 27Y6-15 27Y6-18 2842-17 2858-01 2188-03 26G4-17 2497-22 2538-28 1156-05 1158-10 1461-01 1461-02 1870-06 1870-14 2429-29 1263-19 2071-24 1469-01 1480-18 1925-12 1926-02 1548-17 1845-15 198703 1987-04 1159-20 15D9-20 15D9-19 2145-09 2145-10 2148-16 27Y6-10 27Z0-05 2218-14 27J7-08 2869-19 16Z2-19 2894-14 2137-18 2149-14 2585-24 2384-20 2385-01 27X6-05 2347-09 2347-10 2358-17 1118-11 26C3-14 26C3-20 2708-17 2708-18 2234-03 1156-08 1156-10 1344-09 27Y6-17 27S0-17 27Y6-13 27X4-15 26H4-29 27S2-11 27S2-12 2066-38 1867-20 1868-01 2047-09 2086-35 2369-06 2720-18 16X4-18 1987-14 1874-05 1985-08 1985-10 2494-28 2497-21 2064-06 2067-29 2472-19 2321-21 2195-03 1714-18 26B4-24 26B4-25 2495-02 2543-03 2543-08 2543-09 26E0-22 27Z0-13 2048-32 27S2-02 27S2-03 2704-18 2719-03 1208-08 1208-09 1211-02 1211-03 2367-14 2367-15 2059-10 27S2-04 27S2-08 27R5-02 27R3-18 2051-32 2896-06 2894-13 2058-32 27Y6-12 27S2-13 2102-33 2219-19 2225-01 2234-06 2224-15 2224-16 15C9-09 2229-10 2233-19 2233-13 2234-04 2496-21 1871-03 2382-07 2382-12 2403-21 2493-24 26H3-27 2848-11 26D8-19 26D8-27 2479-08 1601-02 1858-05 2317-18 2332-11 2332-12 2396-13 2398-06 2312-27 2332-20 26D0-24 1853-02 1855-10 1855-11 1858-02 1858-03 1849-02 1853-01 1691-14 1860-12 1860-13 1860-14 1860-11 2104-21 2104-03 2368-04 1128-17 2070-22 2581-19 2581-20 27J6-20 27X4-16 27X4-17 2850-07 2850-08 1146-05 1146-06 1147-04 1485-11 2067-30 2070-40 2414-10 2414-11 2414-12 2427-22 2430-28 2431-24 2431-25 26D9-29 26E4-09 2850-11 28B6-19 1611-19 2192-16 27Z8-10 27Z8-11 26G4-19 2858-05 2860-20 2368-13 2369-13 2304-12 1131-08 1132-01 1132-02 15J6-10 2144-13 2140-04 2894-09 2894-16 2894-17 1983-09 27Z8-15 2822-11 28D7-20 28F5-18 2432-27 1862-27 2719-09 2304-29 2305-09 2566-28 2566-29 2073-09 2059-40 2422-23 2422-24 2149-13 2430-30 2431-21 2431-22 2108-25 2108-26 1988-14 2494-27 2209-09 2209-10 26B6-29 14G5-05 2072-28 27J7-05 26G7-21 2073-40 27Z8-17 2848-13 26F0-30 2534-04 2534-03 2535-09 2535-10 2535-11 2405-28 2405-29 26B3-28 2544-22 2550-18 2488-04 27W5-01 1338-12 1338-13 1338-14 2202-11 2292-17 143115 2209-19 2327-26 1988-10 2056-09 2056-12 1559-13 1539-13 1539-14 1571-12 1480-19 1490-09 1490-13 1490-14 1491-01 15B7-08 28F5-14 2468-17 28E7-15 2432-22 2432-23 190520 2149-12 2104-08 2104-20 1867-16 1156-04 2444-14 1868-02 2423-04 2535-05 2192-13 2196-03 2497-15 2422-16 2137-03 2137-04 1698-05 1711-17 1991-13 1991-14 2233-17 2233-18 2543-11 2543-12 1544-06 26F1-18 26H6-28 26G4-14 2071-23 27Y6-14 2422-28 2423-20 2056-10 2056-13 2581-24 2585-23 2086-36 2108-37 2047-38 0884-03 16L7-18 16P4-01 2085-30 2087-11 26F0-28 26F1-16 27X8-20 15V4-20 2321-22 16P1-12 16P3-02 16P3-03 1865-19 1999-20 1861-20 15H7-19 15H7-20 15H8-01 15H8-09 15H9-13 15H9-14 15H9-15 15H9-16 15J0-03 15J0-04 15J0-06 15J0-08 16K4-01 16K4-02 16K4-03 16K4-04 16K5-07 16K5-08 2267-20 2262-19 2262-20 2221-03 199902 2414-09 2535-08 2385-02 2385-03 2327-12 2867-02 2427-14 2427-15 2234-08 2234-09 2236-10 1008-10 16K6-20 16K7-01 16K7-02 16K7-03 2486-12 2389-16 1142-18 1461-05 1468-18 1468-19 1468-20 2330-16 2330-18 1171-07 1171-08 1171-09 1173-10 1160-08 1171-06 1174-06 1174-07 2275-11 2278-10 2278-11 2330-19 2140-02 2262-17 2267-17 2267-18 2283-17 2330-17 2585-01 2585-02 2585-03 2585-16 2585-17 2399-05 2399-07 2233-15 2891-19 2079-29 15G9-03 15G9-04 2494-29 26G3-30 27J7-12 2305-10 1142-09 1145-10 26E0-20 26E0-21 2293-05 1140-01 2052-05 26D5-23 2196-07 1858-35 26D0-22 26D0-23 2192-09 2322-08 2107-22 2107-23 2190-38 2195-10 2402-29 2195-01 2103-36 2321-29 2066-28 1139-09 1173-08 2070-07 2310-08 1334-15 1335-13 1459-14 14B2-14 2488-03 1428-07 1182-08 1211-06 1331-13 1338-15 1338-16 27S2-14 1147-05 1149-01 26C7-28 14H8-02 1926-04 2432-26 2432-28 2566-27 26E4-13 15J6-02 15J6-08 2427-28 2037-13 2059-12 2058-18 2429-28 1263-18 2148-14 2149-08 2149-20 16P6-19 16R7-03 2256-02 2076-18 2262-16 2414-13 2891-16 2891-17 2148-09 28E7-18 2086-31 0829-20 2894-18 2894-19 1601-01 1698-03 16P6-14 2058-23 1429-13 2250-01 2307-05 1919-02 1919-04 2220-02 2220-04 1469-02 2220-07 26A5-02 2850-14 2431-23 2471-20 27K0-06 1211-20 15B7-06 27R3-09 27R3-13 1907-02 2321-20 2438-04 2409-22 2408-20 15C8-11 14B1-05 14B1-08 2332-10 2438-13 26E0-15 2037-11 1991-17 1991-16 26H6-27 2061-40 0802-01 199010 27Z0-08 15R8-20 27J6-14 27J6-13 27J7-01 27J6-18 27Z8-07 27P2-09 2479-10 2493-17 26D5-21 2850-02 27X6-06 27Y6-09 187018 1925-13 27Z8-06 2488-01 26F1-21 2708-16 27Z8-14 2042-02 2321-27 26A9-10 27L1-12 27L1-13 27L1-03 27L1-04 27K9-16 27L1-01 27L1-02 1983-08 1868-14 2288-08 2328-02 2473-01 2077-36 1928-02 1924-11 2192-11 1149-07 1149-08 1156-03 1167-02 16R0-07 16R0-08 1863-05 27J7-11 27J9-12 27L1-06 1935-16 26G4-08 15C9-08 27L1-07 27L1-08 15C8-12 15C8-13 15C8-14 15C8-15 27L1-10 27L1-11 2301-05 2308-06 2229-03 2229-07 2067-33 27S2-16 27Z8-16 1698-06 1733-17 26D5-27 26E0-17 1862-16 27J6-16 1538-10 27Z2-09 2494-23 2494-24 2494-25 2494-26 28F5-15 28F5-16 16R5-04 16R5-05 2061-34 2064-05 2137-05 2227-12 2227-13 2413-10 2422-15 2226-19 2233-14 1016-05 1017-01 2102-04 2102-05 27R3-14 27R3-16 2578-23 13B8-10 2708-20 27Z6-20 2867-04 2486-22 16L8-08 16M4-14 16M4-15 16M4-16 16M7-04 2550-16 2563-29 1905-06 2848-10 2842-04 2061-37 2061-38 0801-28 0801-29 2413-07 2422-19
Other devices under the same FDA submission
15 devices cleared under K060373, K100853, K953239 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- cobas c 501 Module 4745914001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY, MICRO, FOR CLINICAL USE1,628 reports
- cobas 8000 c 702 Module 6473245001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE620 reports
- cobas e 601 Immunoassay Analyzer 04745922001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE607 reports
- cobas 8000 cobas ISE module 05964075001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE419 reports
- cobas 8000 c 502 Module 5964067001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE403 reports
- cobas c 311 4826876001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY, MICRO, FOR CLINICAL USE203 reports
- cobas 8000 c 701 Module 5641489001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE195 reports
- Sodium electrode 10825468001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, SODIUM99 reports
- Sodium electrode 10825468001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, SODIUM26 reports
- Potassium Electrode 10825441001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, POTASSIUM14 reports
- Chloride Electrode 3246353001Roche Diagnostics GmbH · ELECTRODE, ION-SPECIFIC, CHLORIDE13 reports
- Potassium electrode 10825441001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, POTASSIUM12 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page