Skip to content

BD MAXTM Vaginal Panel. Catalog Number(s): 443712 recallZ-0392-2018 · Becton Dickinson & Co. · initiated Jul 18, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and indeterminate (IND) results for the BD MAXTM Vaginal Panel, experienced by some customers. A non-reportable result may lead to repeat of the test and/or re-collection of a specimen.

FDA root cause category: Use error

What the company is doing

Customers were notified via UPS on 07/18/2017. The recall for BD MAX" Vaginal Panel and BD MAX" UVE Specimen Collection Kit was initiated on July 18, 2017 and expanded to include BD MAX" CT/GC/TV on August 11, 2017. The recall expansion letter to customers was dated 08/14/2017.

Recall details

Recall number
Z-0392-2018
Status
Terminated
Initiated by the company
Jul 18, 2017
Recalling firm
Becton Dickinson & Co.
Product code
NSU
Distribution
US and OUS.

Product description

BD MAXTM Vaginal Panel. Catalog Number(s): 443712 443710 443711 (RUO)for collection and transport of vaginal and endocervical swab specimens, in addition to the preservation of urine from males and females.

Code information (models, lots, serial numbers)

UDI: 30382904433766. All Lot numbers: Not Distributed in US Not Distributed in US

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 78808).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page