Rheumatoid Factor (RF) Kit for use on SPAPLUS, Product… recallZ-0553-2018 · The Binding Site Group, Ltd. · initiated Jan 4, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from The Binding Site Group, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The low and high controls for the product are returning results above the assigned values.

FDA root cause category: Under Investigation by firm

What the company is doing

Binding Site sent an Urgent Field Safety Notice dated December 22, 2017. Advise on action to be taken by the User: " The new QC values detailed below should be used for the remainder of the shelf life of this Rheumatoid Factor Kit for use on SPAPLUS. "Input the revised QC values into your SPAPLUS instrument "Sign and return the TSWS18 E-Back Form accompanying this notification within 5 business days "Refer to Section 9 of the Product Insert SIN282(.A) if QCs are out of range following reassignment. For further questions, please call (858) 453-9177.

Recall details

Recall number
Z-0553-2018
Status
Terminated
Initiated by the company
Jan 4, 2018
Recalling firm
The Binding Site Group, Ltd.
Quantity
49 kits to the U.S. office (32 kits further distributed)
Product code
DHR
FDA submissions
K160070
Distribution
US Distribution was made to CA. There was no foreign/military/government distribution.

Product description

Rheumatoid Factor (RF) Kit for use on SPAPLUS, Product #LK151.S.A Intended for the quantitative in-vitro measurement of rheumatoid factor in serum using the Binding Site SPAPLUS analyser. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis. This test should be used in conjunction with other laboratory and clinical findings

Code information (models, lots, serial numbers)

Lot #416595, Exp. 31/03/2018; UDI #05051700018319