NexGen Legacy recallZ-0663-2018 · Zimmer Orthopedic Mfg Ltd · initiated Jul 24, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Orthopedic Mfg Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a possibility that the protective foam insert used during the packaging of the affected products is undersized, which may cause a breach in the inner cavity during transportation.
FDA root cause category: Reprocessing Controls
What the company is doing
On approximately July 24, 2017, customers were notified via letter of the recall. Instructions for distributors included to ensure affected personnel are aware of the recall, locate and quarantine any affected product in inventory, complete and return the Inventory Return Certification Form, make arrangements to return the product Zimmer Biomet, and provide customers if product was further distributed. Instructions for Risk Managers at medical facilities include to ensure affected personnel are aware of the recall, to assist the Zimmer Biomet sales representative in identifying and quarantining affected devices so they can be removed by the sales representative, and to complete and return the Certification of Acknowledgement form. Further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. or 1 (800) 613-6131
Recall details
- Recall number
- Z-0663-2018
- Status
- Terminated
- Initiated by the company
- Jul 24, 2017
- Recalling firm
- Zimmer Orthopedic Mfg Ltd
- Quantity
- 7 devices
- Product code
- NJL
- FDA submissions
- P060037
- Distribution
- Worldwide Distribution - US Distribution to the states of : CA, MA, ME, MI, MN and WI., and to the countries of Germany, France and Italy.
Product description
NexGen LPS Femoral Component, Left and Right, Size G Replaces the femoral condyle of the knee joint in Total Knee Arthroplasty.
Code information (models, lots, serial numbers)
Item Number: 00-5964-017-51, Lot Numbers (UDI Number): 63329529 ((01) 00889024001152 (17) 260430 (10) 63329529); 63342472 ((01) 00889024001152 (17) 260430 (10) 63342472); 63329533 ((01) 00889024001152 (17) 260430 (10) 63329533); 63329527 ((01) 00889024001152 (17) 260430 (10) 63329527)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NexGen® Legacy® 5964-17-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing1 recall3 reports
Product lines
Other devices under the same FDA submission
25 devices cleared under P060037 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NexGen® Legacy® 5964-15-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing7 reports
- NexGen® Legacy® 5964-16-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing7 reports
- NexGen® Legacy® 5964-16-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing7 reports
- NexGen® Legacy® 5964-17-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing6 reports
- NexGen® Legacy® 5964-15-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing4 reports
- NexGen® Legacy® 5996-16-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing3 reports
- NexGen® Legacy® 5996-17-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing3 reports
- NexGen® Legacy® 5996-17-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing3 reports
- NexGen® Legacy® Prolong® 00-5917-050-10Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing2 reports
- NexGen® Legacy® 5960-15-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing2 reports
- NexGen® Legacy® 5964-13-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing2 reports
- NexGen® Legacy® 5964-14-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page