ZIPTIGHT FIXATION SYSTEM recallZ-0818-2018 · Zimmer Biomet, Inc. · initiated Aug 22, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Ziptight Ankle Syndesmosis Fixation was packaged without the tip protectors.
FDA root cause category: Packaging process control
What the company is doing
Customers were notified via letter on about 08/22/2017. Instructions for distributors, sales representatives, and distributor operation managers include to ensure affected personnel are aware of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return affected product, and identify any additional hospitals or surgeons that should be notified. Instructions for surgeons and risk managers include to ensure affected personnel are aware of the recall, identify and quarantine any affected product on hand, complete and return the Certificate of Acknowledgement, return any affected product. Questions or concerns can be addressed to the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST, Monday through Friday. Additionally, questions may be sent by email to [email protected].
Recall details
- Recall number
- Z-0818-2018
- Status
- Terminated
- Initiated by the company
- Aug 22, 2017
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 885 units total
- Product code
- MBI
- FDA submissions
- K083070, K130033
- Distribution
- Distributed domestically to AK, AL, AR, AZ, CA, CO, FL, ID, IN, KY, MA, MD, ME, MI, MN, MO, NC, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI. Distributed internationally to ARGENTINA, AUSTRALIA, CANADA, COSTA RICA, JAPAN, NETHERLANDS.
Product description
Ziptight Ankle Syndesmosis Fixation Device, Titanium Implant Titanium Implant for surgical repairs of soft tissue.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Item No. 905759, Lot No. (UDI No.): 58860 ((01)00880304476103(17)220523(10)058860), 58880 ((01)00880304476103(17)220608(10)058880), 58890 ((01)00880304476103(17)220608(10)058890), 58910 ((01)00880304476103(17)220608(10)058910), 116030 ((01)00880304476103(17)220519(10)116030), 169780 ((01)00880304476103(17)220609(10)169780), 169790 ((01)00880304476103(17)220608(10)169790), 213450 ((01)00880304476103(17)220609(10)213450), 213480 ((01)00880304476103(17)220607(10)213480), 374420 ((01)00880304476103(17)220319(10)374420), 392530 ((01)00880304476103(17)220610(10)392530), 476670 ((01)00880304476103(17)220607(10)476670), 503830 ((01)00880304476103(17)220608(10)503830), 614910 ((01)00880304476103(17)220421(10)614910), 670640((01)00880304476103(17)220621(10)670640), 728190 ((01)00880304476103(17)220606(10)728190), 984070 ((01)00880304476103(17)220512(10)984070), 984080 ((01)00880304476103(17)220607(10)984080)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ZIPTIGHT FIXATION SYSTEM 904759Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall23 reports
Product lines
Other devices under the same FDA submission
11 devices cleared under K083070, K130033 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ZIPTIGHT 904834Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall9 reports
- ZIPTIGHT FIXATION SYSTEM 909853Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE15 reports
- TOGGLELOC XL WITH ZIPLOOP INLINE 110005090Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE4 reports
- ToggleLoc With Ziploop 904755Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE3 reports
- ToggleLoc With Ziploop 904754Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE3 reports
- ZIPTIGHT FIXATION SYSTEM 909856Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE3 reports
- TOGGLELOC XL WITH ZIPLOOP 110005089Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE2 reports
- ToggleLoc With Ziploop 904756Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE1 report
- ToggleLoc with Ziploop 909854Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
- ZIPTIGHT FIXATION SYSTEM 909857Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
- ZIPTIGHT 904835Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page