Skip to content

Flex-Neck recallZ-1143-2018 · Merit Medical Systems, Inc. · initiated Sep 20, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

2 Cuff catheters were incorrectly configured and labeled as 1 Cuff product.

FDA root cause category: Mixed-up of materials/components

What the company is doing

On 09/20/17, an URGENT PRODUCT RECALL NOTICE was sent to all direct consignees via mail, informing them of a voluntary recall of specific lots of peritoneal dialysis catheters and catheter containing kits. Consignees were advised to immediately do the following: identify and quarantine effected devices; inform personnel to whom the devices were distributed of the field action; fill out and email the customer response form to [email protected] within 5 days, and return effected devices within ten days using Federal Express account 112952756. Consignee questions were directed to phone: 801-208-4381 or fax: 801-316-4880. Sales representatives were sent an email communication informing them of the recall and instructing them to contact effected customers to help them return Customer Response Forms and product.

Recall details

Recall number
Z-1143-2018
Status
Terminated
Initiated by the company
Sep 20, 2017
Recalling firm
Merit Medical Systems, Inc.
Quantity
30
Product code
FJS
FDA submissions
K031351
Distribution
United States, Austria, Finland, Germany, Ireland, Switzerland, and United Kingdom

Product description

Flex-Neck Classic Peritoneal Dialysis Catheter 1 Cuff Infant Coiled, Catalog # CF-4235/B

Code information (models, lots, serial numbers)

Lot #: a) H1136644, Exp 12/31/2018, UDI (01)00884450305850(17)181231(10)H1136644; b) H1139640, Exp 12/31/18, UDI (01)00884450305850(17)181231(10)H1139640

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 79426).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page