Oncomine Dx Target Test User Guides and Assay Definition… recallZ-1279-2018 · Life Technologies Corporation · initiated Dec 4, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Life Technologies Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Oncomine Dx Target Test may report erroneous results when used off-label for samples other than non-small cell lung cancer (NSCLC) tumor specimens.

FDA root cause category: Device Design

What the company is doing

The following actions are described in the consignee later: Review your testing results from the Oncomine Dx Target Test to ensure that none of the reports containing BRAF V600E positive samples are impacted. Utilize Figure 1 and 2 to aid in identifying this discrepancy. If you have any questions, please contact 408.802.2341.

Recall details

Recall number
Z-1279-2018
Status
Terminated
Initiated by the company
Dec 4, 2017
Recalling firm
Life Technologies Corporation
Quantity
9
Product code
PQP
FDA submissions
P160045
Distribution
US Distribution to the states of: AZ, CA, NC, TX, and VA.

Product description

Oncomine Dx Target Test User Guides and Assay Definition File, Model: A32461; UDI: (01)10190302006071(11)170911(10)1.8(240)A32461 The Oncomine Dx Target Test is a qualitative in vitro diagnostic test that uses targeted high- throughput, parallel-sequencing technology to detect single-nucleotide variants (SNVs) and deletions in 23 genes from DNA and fusions in ROS1 from RNA isolated from formalin-fixed, paraffin-embedded (FFPE) tumor tissue samples from patients with non-small cell lung cancer (NSCLC) using the Ion PGM Dx System. The test is indicated to aid in selecting NSCLC patients for treatment with the targeted therapies listed in Table 1 in accordance with the approved therapeutic product labeling.

Code information (models, lots, serial numbers)

Lot Numbers: v1.8 (Rev. C) (REVC20170906) and v1.8 (Rev. C) (REVC27JUN2017)  UDI: (01)10190302006071

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

5 devices cleared under P160045 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.