Skip to content

ARCHITECT recallZ-1629-2018 · Abbott Gmbh & Co. KG · initiated Dec 14, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Gmbh & Co. KG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Falsely elevated results may be obtained when using the ARCHITECT DHEA-S assay with samples from infants up to 60 days old.

FDA root cause category: Device Design

What the company is doing

Customers were notified via letter on about 12/14/2017. Instructions included to immediately discontinue use of the ARCHITECH DHEA-S assay with samples form infants up to 60 days old, to immediately discontinue use of the ARCHITECH DHEA-S reagent insert Expected Values representative data for children up to 10 years old, to review the letter with the Medial Director and follow the laboratory protocol regarding the need for reviewing previously reported patient results, to forward the Product Correction letter to customers if the product was further distributed, and to complete and return the Customer Reply Form. Questions from U.S. customers can be directed to Customer Service at 1-877-4ABBOTT (available 24 hours per day, 7 days a week) and customers outside the U.S. should contact their local area Customer Service.

Recall details

Recall number
Z-1629-2018
Status
Terminated
Initiated by the company
Dec 14, 2017
Recalling firm
Abbott Gmbh & Co. KG
Quantity
10,153 units
Product code
JKC
Distribution
Worldwide Distribution in the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MN, MO, MS, NC, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, VT, WA, WY. Distributed internationally to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Barbados, Belarus, Belgium, Bosnia and Herzegovi, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Egypt, Finland, France, Gaza and Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mexico, Morocco, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Yemen, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Tajikistan, Thailand, Tunisia, Turkey, Uganda, United Arab Emirates, Uganda, United Kingdom, Uruguay .

Product description

ARCHITECT DHEA-S Reagent Product Usage: It is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of dehydroepiandrosterone sulfate (DHEA-S) in human serum and plasma on the ARCHITECT iSystem.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

List No. 8K2-20 with the following Lot No., Exp. Date, UDI:  a. Lot No. 01316K000, Exp. Date 04FEB2018, UDI (01) 00380740017446 (17) 180204 (10) 01316K000 (240) 8K2720;  b. Lot No. 01217A000, Exp. Date 13APR2018, UDI (01) 00380740017446 (17) 180413 (10) 01217A000 (240) 8K2720;  c. Lot No. 02217D000, Exp. Date 24JUN2018, UDI (01) 00380740017446 (17) 180624 (10) 02217D000 (240) 8K2720;  d. Lot No. 04117H000, Exp. Date 25NOV2018, UDI (01) 00380740017446 (17) 181125 (10) 04117H000 (240) 8K2720;   List No. 8K27-25 with the following Lot No., Exp. Date, UDI:  a. Lot No. 01216K000, Exp. Date 04FEB2018, UDI (01) 00380740017453 (17) 180204 (10) 01216K000 (240) 8K2725;  b. Lot No. 01117A000, Exp. Date 13APR2018, UDI (01) 00380740017453 (17) 180413 (10) 01117A000 (240) 8K2725;  c. Lot No. 02217C000, Exp. Date 17JUN2018, UDI (01) 00380740017453 (17) 180617 (10) 02217C000 (240) 8K2725;  d. Lot No. 02317D000, Exp. Date 24JUN2018, UDI (01) 00380740017453 (17) 180624 (10) 02317D000 (240) 8K2725;  e. Lot No. 02417D000, Exp. Date 24JUN2018, UDI (01) 00380740017453 (17) 180624 (10) 02417D000 (240) 8K2725;  f. Lot No. 00517F000, Exp. Date 02SEP2018, UDI (01) 00380740017453 (17) 180902 (10) 00517F000 (240) 8K2725;  g. Lot No. 02817H000, Exp. Date 21OCT2018, UDI (01) 00380740017453 (17) 181021 (10) 02817H000 (240) 8K2725

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page