Skip to content

CONNECTOR recallZ-1363-2018 · Orthofix, Inc · initiated Feb 23, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Orthofix, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The firm disseminated the recall notices on 02/23/2018 by mail. The firm requested the return of the screws.

Recall details

Recall number
Z-1363-2018
Status
Terminated
Initiated by the company
Feb 23, 2018
Recalling firm
Orthofix, Inc
Quantity
397 units
Product code
NKG
FDA submissions
K172194
Distribution
US

Product description

ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003

Code information (models, lots, serial numbers)

UDI 18257200125371, Lot Numbers: O02, O03 and O04

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 79498).

Sources and updates