Skip to content

Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref… recallZ-1301-2018 · The Binding Site Group, Ltd. · initiated Mar 9, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from The Binding Site Group, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The product is not meeting the antigen excess protection claim of up to 760mg/L, as stated in the Instructions for Use.

FDA root cause category: Under Investigation by firm

What the company is doing

The recalling firm notified their U.S. location by letter dated 3/6/2018 via email on 3/9/2018.

Recall details

Recall number
Z-1301-2018
Status
Terminated
Initiated by the company
Mar 9, 2018
Recalling firm
The Binding Site Group, Ltd.
Quantity
20 kits distributed to the U.S.
Product code
JZG
FDA submissions
K122066
Distribution
US Distribution was made to CA. There was no foreign/military/government distribution.

Product description

Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref LK043.L.OPT.A, IVD. The firm name on the label is The Binding Site Group Ltd., Birmingham, U.K. The Optilite Beta-2-Microglobulin (2M) Urine kit is intended for the quantitative in-vitro measurement of beta-2-microglobulin in urine using the Binding Site Optilite analyser to aid in the diagnosis of active rheumatoid arthritis and kidney disease. This test should be used in conjunction with other laboratory and clinical findings.

Code information (models, lots, serial numbers)

Lot 407460, UDI 05051700017688

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page