Oncomine Dx Target Test recallZ-1480-2018 · Life Technologies Corporation · initiated Feb 8, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Life Technologies Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The lower concentration of the RNA panel could cause a NO CALL or FALSE NEGATIVE for the reporting of ROS1 fusion. If a physician relies on a false negative result, crizotinib could be improperly excluded from the patient's treatment regimen.

FDA root cause category: Process control

What the company is doing

The firm issued an Urgent: Medical Device Recall Notification, dated February 16, 2018, via e-mail and phone call to affected consignees. The letter identified the affected device and advised customers of the potential "No Call" or false positive outcomes. Users were advised that they may need an alternate method to determine the presence of ROS1 variant in previously tested patient samples, and that any future tests using the affected lot cannot be considered valid if "No Call" or "Negative" is the result. Requested action included reviewing previous results which used the affected lot. Customers are to complete the Customer Response Sheet provided and e-mail a scanned copy along with the letter to [email protected]. Questions and comments should be directed to Vineet Bansal, 408-802-2341, [email protected].

Recall details

Recall number
Z-1480-2018
Status
Terminated
Initiated by the company
Feb 8, 2018
Recalling firm
Life Technologies Corporation
Quantity
30
Product code
PQP
FDA submissions
P160045
Distribution
Distributed to states: AZ, CA, NC, TX, and VA.

Product description

Oncomine Dx Target Test, (Model No. A32451), Oncomine Dx Target RNA/DNA Panel (Model No. A32441). A qualitative In Vitro Diagnostic test.

Code information (models, lots, serial numbers)

Lot 1705001 UDI: (01)10190302006385(17)180220(10)1705001(240)A32441

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

5 devices cleared under P160045 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page