Skip to content

Analyzer and Atellica IM 1600 Analyzer system for in vitro… recallZ-1588-2018 · Siemens Healthcare Diagnostics, Inc. · initiated Feb 26, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers

FDA root cause category: Software design

What the company is doing

Recall initiated 02/23/2018 Firm released an update software (SW) version (V), 1.12.1, which corrects the behavior. A Mandatory Update (MU) campaign was launched to install the new SW V on all systems, whether or not they are impacted by this behavior. The MU campaign will be tracked to 100% completion globally.

Recall details

Recall number
Z-1588-2018
Status
Terminated
Initiated by the company
Feb 26, 2018
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product code
JJE
FDA submissions
K151792
Distribution
FL and France Spain Sweden United Kingdom Germany

Product description

Analyzer and Atellica IM 1600 Analyzer system for in vitro diagnostic testing of clinical specimens.

Code information (models, lots, serial numbers)

UDI 00630414002026

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K151792 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 79683).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page