LINX Reflux Management System, Implant Clasp, 13-16 Bead… recallZ-2041-2018 · Torax Medical, Inc. · initiated Apr 20, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Torax Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possible out of specification condition which may allow a bead component to separate from an adjacent wire link. This condition may result in a discontinuous or open LINX device.
FDA root cause category: Nonconforming Material/Component
What the company is doing
The firm, Consignees Torax Medical Inc., sent an "URGENT: FIELD SAFETY NOTICE (REMOVAL)" letter mailed on approximately 04/20/2018 (US consignees and physicians) and 04/23/2018 (OUS consignees and physicians). The letter described the product, problem and actions to be taken. The customers were instructed to examine inventory for affected product, quarantine any affected product, communicate the issue to all relevant operating room or materials management personnel, forward the recall information to facilities where the product may have been forwarded to, complete the Business Reply Form confirming receipt of this notice and return it to Torax Medical, Inc. within three (3) business days via email at [email protected]; ( return the BRF even if you do not have the product lot subject to this recall.), and arrange the return of any affected product on hand. Additionally, a recall notification will be sent to sent to physicians outlining the reason for recall and to advise them to monitor their patients. Question regarding the recall can be directed to the Customer Support Cetner at 1-877-384-4266, which is open Monday through Friday, 7:30 AM to 6:30 PM ET.
Recall details
- Recall number
- Z-2041-2018
- Status
- Terminated
- Initiated by the company
- Apr 20, 2018
- Recalling firm
- Torax Medical, Inc.
- Quantity
- 9,131 devices in total
- Product code
- LEI
- FDA submissions
- P100049
- Distribution
- Worldwide Distribution: US (nationwide) in states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY; and in countries of: Austria, France, Germany, Italy, Scotland, Spain, Switzerland, United Kingdom.
Product description
LINX Reflux Management System, Implant Clasp, 13-16 Bead, 1.5T, Product Code LXM-xx, sterile, Rx only. Labeled as: a. Product Code LXM-13; b. Product Code LXM-14; c. Product Code LXM-15; d. Product Code LXM-16; Laparoscopic, fundic-sparing anti-reflux procedure indicated for patients diagnosed with Gastroesophageal Reflux Disease (GERD) as defined by abnormal pH testing, and who are seeking an alternative to continuous acid suppression therapy (i.e. proton pump inhibitors or equivalent) in the management of their GERD.
Code information (models, lots, serial numbers)
a. Product Code LXM-13, Part No. 4549-13, GTIN 00855106005394; b. Product Code LXM-14, Part No. 4549-14, GTIN 00855106005400; c. Product Code LXM-15, Part No. 4549-15, GTIN 00855106005417; d. Product Code LXM-16, Part No. 4549-16, GTIN 00855106005424; Lot No. 6100 through 14055, 15286, 15287, 15314, 15315, 15317, and 16557
Other devices under the same FDA submission
18 devices cleared under P100049 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LINX LXMC14Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall289 reports
- LINX LXMC15Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall261 reports
- LINX LXMC13Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall169 reports
- LINX LXMC16Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall167 reports
- LINX LXMC17Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall145 reports
- LINX LS-[13]Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall28 reports
- LINX LS-[14]Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall24 reports
- LINX LXC-[13]Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall24 reports
- LINX LXC-[14]Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall18 reports
- LINX LXC-[15]Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall16 reports
- LINX LXC-[12]Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall12 reports
- LINX LXC-[16]Torax Medical, Inc. · Implant, Anti-Gastroesophageal Reflux1 recall11 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 79822).
- Z-2037-2018 · TerminatedLINX Reflux Management System, Implant Suture, 12-17 Bead, 0.7T, Product Code LS-xx, sterile, Rx only. Labeled as: a. Product Code LS-12; b. Product Code LS-13; c. Product Cod…
- Z-2038-2018 · TerminatedLINX Reflux Management System, Implant Clasp, 12-17 Bead, 0.7T, Product Code LXC-xx, sterile, Rx only. Labeled as: a. Product Code LXC-12; b. Product Code LXC-13; c. Product C…
- Z-2039-2018 · TerminatedLINX Reflux Management System, Implant Clasp, 13-17 Bead, 1.5T, Product Code LXMC-xx, sterile, Rx only. Labeled as: a. Product Code LXMC-13; b. Product Code LXMC-14; c. Produc…
- Z-2040-2018 · TerminatedLINX Reflux Management System, Implant Clasp, 12-17 Bead, 0.7T, Product Code LX-xx, sterile, Rx only. Labeled as: a. Product Code LX-12; b. Product Code LX-13; c. Product Code…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page