Skip to content

Wet-Field Osher Thermodot Marker, 25ga Fine Tip, Curved recallZ-1539-2018 · Beaver-Visitec International Inc. · initiated Feb 26, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beaver-Visitec International Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect tip was used during manufacturing of the Osher Thermodot Marker. The markers have a needle tip instead of the proper flat tip.

FDA root cause category: Process control

What the company is doing

On February 26, 2018, Beaver Visitec International Inc. emailed Product Recall Notices to affected customers. The letter identified the affected device and the reason for the recall. Customers were instructed to destroy affected product, provide evidence of the destruction, and complete a customer response form. Customers will receive a credit. Customers with questions regarding the recall may contact Customer Service at (866)906-8080.

Recall details

Recall number
Z-1539-2018
Status
Terminated
Initiated by the company
Feb 26, 2018
Recalling firm
Beaver-Visitec International Inc.
Quantity
110 total products
Product code
GEI
FDA submissions
K911160
Distribution
Distributed to the states of OH, NY, FL, MN, and KS.

Product description

WET-FIELD OSHER THERMODOT MARKER, 25GA FINE TIP, CURVED; Part number 221270. Designed to create an ink-free, fine dot on the eye that can be used as a precise reference point for a variety of procedures.

Code information (models, lots, serial numbers)

Lot numbers: 6000292 exp. 7/25/20 and 6000540 exp. 08/30/20  UDI: 30886158014899

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates