Why it was recalled

Urea Nitrogen reagent kit demonstrated an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails.

FDA root cause category: Material/Component Contamination

What the company is doing

Date recall was initiated: 03-27-18 Customer communication was sent on 04-03-18 Urgent Medical Device Recall (CHC18-04.A.US) was distributed to all customers in the United States and an Urgent Field Safety Notice (CHC18-04.A.OUS) was distributed to all customers outside the United States who have received ADVIA Chemistry Urea Nitrogen reagent kit lots 408143 and 418290 (Lot 418290 was not sold in the US). These letters instruct the customers to discontinue use of reagent kit lots 408143 and 418290 and to contact their Customer Care Center for replacements kits. In the US the method of distribution is Fed Ex and outside the US the distribution is determined at the country level.

Recall details

Recall number
Z-2062-2018
Status
Terminated
Initiated by the company
Mar 27, 2018
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product code
CDQ
FDA submissions
K991576
Distribution
Nationwide and worldwide

Product description

ADVIA Chemistry Urea Nitrogen Reagent (material number 10309051) for use on ADVIA 1800, 2400 and XPT Chemistry Systems

Code information (models, lots, serial numbers)

Lot # Expiry UDI# 408143 2018-06-28 00630414517513 418290 2018-06 - 28 00630414517513

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

6 devices cleared under K991576 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page