EXACTAMIX recallZ-1909-2018 · Baxter Healthcare Corporation · initiated Apr 25, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific lots of EXACTAMIX Inlets listed below due to incorrectly performed over-labeling. The inlets were over-labeled with a blank label, or with a label containing an extended expiration date. If the product is used beyond the expiration date,
FDA root cause category: Labeling Change Control
What the company is doing
The firm initiated the recall by letter on 04/25/2018. The letter identified the affected product, problem and actions to be taken. the firm requested the return of the product. Distributors were directed to conduct a sub-recall, requesting their customers return the product to the entity from whom they received the product.
Recall details
- Recall number
- Z-1909-2018
- Status
- Terminated
- Initiated by the company
- Apr 25, 2018
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 59475 units
- Product code
- LHI
- FDA submissions
- K002705
- Distribution
- Worldwide Distribution - US Distribution
Product description
Baxter EXACTAMIX Inlet, Vented, High-Volume Inlet, REF H938174 Usage: Exacta-Mix 2400 Compounding System Administration set is a disposable component of a compounding device used in the pharmacy to compound multiple source ingredients into one final solution.
Code information (models, lots, serial numbers)
UDI: 00085412475790 Lot Numbers: 802279, 802280, 802281
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EXACTAMIX H938174Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING2 recalls238 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K002705 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EXACTAMIX H938724Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall513 reports
- EXACTAMIX H938175Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING3 recalls263 reports
- EXACTAMIX H938173Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls260 reports
- EXACTAMIX H938176Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls232 reports
- EXACTAMIX H938724EBaxter Healthcare Corporation · Set, i.V. Fluid transfer105 reports
- EXACTAMIX H938792Baxter Healthcare Corporation · Set, i.V. Fluid transfer12 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 79975).
- Z-1908-2018 · TerminatedBaxter EXACTAMIX Inlet, Non-Vented, High-Volume Inlet, REF H938173 Product Usage: Exacta-Mix 2400 Compounding System Administration set is a disposable component of a compounding …
- Z-1910-2018 · TerminatedBaxter EXACTAMIX Inlet, Vented, Micro-Volume Inlet, REF H938175 Usage: Exacta-Mix 2400 Compounding System Administration set is a disposable component of a compounding device used…
- Z-1911-2018 · TerminatedBaxter EXACTAMIX Inlet, Syringe Inlet, REF H938176 Usage: Exacta-Mix 2400 Compounding System Administration set is a disposable component of a compounding device used in the pharm…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page