Skip to content

Aortic Arterial Cannula 20 Fr (O.D.) recallZ-2159-2018 · Maquet Cardiopulmonary Ag · initiated May 14, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Cardiopulmonary Ag or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

This lot of arterial cannula may have been supplied with 3/8-width de-airing plugs that could break during the process of removal from the arterial cannula.

FDA root cause category: Component design/selection

What the company is doing

Urgent Medical Device Recall letters were distributed beginning 5/14/18. The letters instructed customers to do the following: 1. Read and understand this notification, and disseminate the information provided to all users, or prospective users, of the affected arterial cannulae. 2. Check your arterial cannula inventory and compare with the Getinge part number, article number and lot number provided on Page 1 of this notification. 3. Remove any affected product that are still in your inventory and prepare it for shipment back to Getinge. 4. Call Getinge Customer Support at 1-888-627-8383 (Press Option 2 and then Option 2) for a Return Authorization (RGA) and for shipping information and label. Customer Support will issue a credit for the product return. 5. Complete, sign and promptly return the filled out response form (Page 4 of this notification) either by email to [email protected] or by fax to 1-973-860-1675.

Recall details

Recall number
Z-2159-2018
Status
Terminated
Initiated by the company
May 14, 2018
Recalling firm
Maquet Cardiopulmonary Ag
Quantity
40 US
Product code
DWF
FDA submissions
K012774
Distribution
Worldwide distribution including US state of VA and Puerto Rico, Australia, Belgium, Brazil, Canada, Chile, Colombia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Iran, Ireland, Israel, Italy, Japan, Lebanon, Malaysia, Martinique, Mauritius, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, United Arab Emirates, and United Kingdom.

Product description

Aortic Arterial Cannula 20 Fr (O.D.); 23 cm (L); with 3/8-inch wide connector, Part Number 701002206, Article Number A20-0101. For use in cardiopulmonary bypass

Code information (models, lots, serial numbers)

UDI CODE: 04037691085555, Lot Number: 92196585

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 80076).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page