Invia Motion Negative Pressure Wound Therapy System.… recallZ-2528-2018 · Medela Inc · initiated Feb 7, 2018
Class II Terminated All recalls posted in July 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medela Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Device may display a battery missing error.
FDA root cause category: Device Design
What the company is doing
Medela sent an Notification letter dated February 1, 2018 to customers via Certified Mail. The letter identified the affected product, problem and actions to be taken. Instructions included to place all affected items on hold until an insert outlining additional charging instructions were included with the devices. Those instructions included to provide a printed copy of the charging instructions in each device package, review those insert instructions with patients and care-givers, and complete and return the confirmation letter. For questions contact Medela Partner Services at [email protected] or via phone 877-735-1626.
Recall details
- Recall number
- Z-2528-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 7, 2018
- Posted by the FDA
- Jul 23, 2018
- Recalling firm
- Medela Inc
- Quantity
- 778 units
- Product code
- OMP
- FDA submissions
- K161128
- Distribution
- US Nationwide Distribution domestically to AR, CA, FL, GA, IL, IN, LA, MI, MN, MO, NC, NJ, NY, OH, OK, PA, TX. Distributed internationally to Spain, Italy, Denmark, Sweden, Switzerland, Netherlands.
Product description
Invia Motion Negative Pressure Wound Therapy System. Labeled as: a. Endure Negative Pressure Wound Therapy (Model 087.4010); b. 15 Days Negative Pressure Wound Therapy (Model 087.4012); c. 60 Days Negative Pressure Wound Therapy (Model 087.4014); d. Endure Negative Pressure Wound Therapy (Model 087.4020 ); e. 15 Days Negative Pressure Wound Therapy (Model 087.4022); f. 60 Days Negative Pressure Wound Therapy (Model 087.4024). Product Usage: The Invia Motion Negative Pressure Wound Therapy (NPWT) system is indicated for patients who would benefit from a suction device (NPWT) as it creates an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
a. Endure Negative Pressure Wound Therapy (Expected life 3 years); b. 15 Days Negative Pressure Wound Therapy (Expected life 15 days); c. 60 Days Negative Pressure Wound Therapy (Expected life 60 days); d. Endure Negative Pressure Wound Therapy (Expected life 3 years); e. 15 Days Negative Pressure Wound Therapy (Expected life 15 days); f. 60 Days Negative Pressure Wound Therapy (Expected life 60 days). Serial No. 1670458 1670488; 1670490; 1670492; 1670496 1670512; 1675368 1675419; 1675421 1675431; 1675433 1675443; 1675449; 1676316 1676365; 1676662 1676709; 1676711; 1676713; 1676715; 1676718 - 1676753; 1676755 1676756; 1676758 1676759; 1676762 1676861; 1677718 1677767; 1679752 1679765; 1679767 1679851; 1689658 1689757; 1696087 1696088; 1696095 1696111; 1696113 1696136; 1696356 1696455; 1697916 1697965; 160884 885; 1660887; 1660894 895; 1660901 903; 1660907; 1660915; 1660918 919; 1660921 22; 1660929; 1660934 935; 1660937; 1660940; 1660953; 1660955; 1660957 959; 1660961 963; 1660966; 1660969; 1660977 980; 1660986; 1660988; 1660991 998; 16611001 002; 1661005; 1661013 014; 1661018; 1661020; 1661032; 1661042; 1661052; 1661055; 1661062; 1661064; 1661066; 1661069; 1661076; 1661078; 1661081; 1661083; 1661086; 1661090; 1661092; 1661094 096; 1661099 112; 1661115; 1661117; 1661122; 1661123; 1662223; 1662230; 1662233; 1662237; 1662241 243; 1662245; 1662247 248; 1662252; 1662255 260; 1662263; 1662265 266; 1668362 369; 1668380 385; 1668392 394; 1668396 398; 1668402 403; 1668426; 1668430; 1668437; 1668441; 1668447; 1668451 452; 1668454 455; 1668458; 1668463; 1668466; 1668469; 1668473; 1668488; 1668494 495; 1668506 507; 1668512 515; 1668520 529; 1669591 593; 1669595 601; 1670635; 1670638; 1670641 642; 1670683 689; 1670694; 1670696; 1670698 706; 1670997; 1660883 884; 1660886; 1660888 893; 1660895 900; 1660904 906; 1660908 914; 1660916 917; 1660920; 1660923 933; 1660936 952; 1660954; 1660956; 1660960; 1660964 965; 1660967; 1660974 976; 1660981 984; 1660986 987; 1661000; 1661010; 1661021; 1661029; 1661036; 1661040; 1661049; 1661056; 1661060; 1661065; 1661073; 1661080; 1661084 085; 1661087; 1661089; 1661091; 1661093; 1661095; 1661097 098; 1661100 116; 1661118 121; 1661123 124; 1662225; 1662227; 1662232; 1662234; 1662236; 1662238 240; 1662244; 1662249 250; 1662253 254; 1662261 262; 1662264; 1668350; 1668375 376; 1668379; 1668386 387; 1668389 391; 1668395; 1668399 401; 1668405; 1668407 408; 1668410 416; 1668418 419; 1668421 425; 1668427; 1668432 439; 1668443 445; 1668447; 1668449; 1668453; 1668456 457; 1668459 462; 1668467 468; 1668471; 1668474; 1668481; 1668484; 1668492 494; 1668496 497; 1668501 502; 1668509 510; 1668514; 1668516 517; 1668520; 1668522;
Other devices under the same FDA submission
2 devices cleared under K161128 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Invia Motion Endure 200.8587Medela AG · Negative Pressure Wound Therapy Powered Suction Pump7 reports
- Invia Motion 15 Days 200.8589Medela AG · Negative Pressure Wound Therapy Powered Suction Pump2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page