Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-T 4906… recallZ-2179-2018 · Datascope Corporation · initiated Jul 27, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for the over-pressure safety (OPS) valve in the tubing kits to not allow flow through.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 7/27/17, an Urgent Safety Alert was issued to customers of the Cardiovascular Procedure Kit with LN130B Over-pressure Safety Valves. The safety alert instructed customers to perform the following: Please examine your inventory immediately to determine if you have unused kits of the batch numbers listed. If so, please follow the recommended actions outlined in the attached Terumo Cardiovascular Systems Urgent Safety Alert. It is possible that custom tubing kits with the affected valves may have already been used without issue. Customers were instructed to test the valve prior to use. The notice was sent to customers again on 10/3/17 and 12/12/17. On 1/11/18, Urgent Medical Device Recall letters were sent to customers. The letters included the following instructions: Terumo CVS now recommends removal of any remaining product from your inventory. A review of our records indicates that you have received the following Maquet/Getinge Custom Tubing Kit batch numbers affected by the Terumo Cardiovascular Systems Medical Device Recall (refer to information provided in the table below). Please examine your inventory immediately to determine if you have unused custom tubing kits of the batch numbers listed. If so, please remove the affected products, quarantine them and follow the instructions for returning the product to Maquet/Getinge for full credit. It is possible that custom tubing kits with the affected valves may have already been used without issue. The Terumo Cardiovascular Systems Medical Device Recall affects the Custom Tubing Kit with LN130B Over-pressure Safety Valve distributed by Maquet/Getinge having lot/batch numbers listed above; no other product lot/batch numbers that were delivered to you are affected. Instructions for returning affected product: 1. Review this notice and Terumos Medical Device Recall Notice (attached). 2. Ensure that all users (i.e. perfusionist, operating room staff and biomedical staff) receive notice
Recall details
- Recall number
- Z-2179-2018
- Status
- Terminated
- Initiated by the company
- Jul 27, 2017
- Recalling firm
- Datascope Corporation
- Quantity
- 27
- Product code
- DWE
- FDA submissions
- K080592
- Distribution
- The products were distributed to the following US states: IA and FL.
Product description
Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-T 4906 PEDIARIC CPB PACK, Part Number 701056041 Product Usage: The Maquet Cardiopulmonary Bypass Custom Tubing kit is a sterile, non-pyrogenic device for single use only and is not to be re-sterilized by the user. The product consists of tubing and connectors with connected oxygenators, reservoirs, filters and other cardiopulmonary bypass components assembled into user specified circuits for procedures requiring extracorporeal support for up to six hours. See Attachment 6 for a copy of MCP Instructions for Use.
Code information (models, lots, serial numbers)
UDI 0607567203600
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Pediatric CPB Pack BEQ-T 4906Datascope Corp. · Tubing, Pump, Cardiopulmonary Bypass1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 80081).
- Z-2180-2018 · TerminatedMaquet Cardiopulmonary Bypass Custom Tubing kit BO-TOP 38600 ADULT CPB, Part Number 701056035R01 Product Usage: The Maquet Cardiopulmonary Bypass Custom Tubing kit is a sterile, n…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page