Ankle Arthrodesis Nail, right T2 Ankle 10x300mm, Product… recallZ-2407-2018 · Stryker GmbH · initiated Apr 27, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
FDA root cause category: Employee error
What the company is doing
Urgent Product Recall letters dated 4/27/18 were distributed to customers. The letters instructed customers to perform the following actions: 1. Inform individuals within your organization who need to be aware of this device recall. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. Response is required, even though you may not have any physical inventory on site anymore. 3. Quarantine and discontinue use of the recalled devices. 4. Maintain awareness of this notice internally until all required actions have been completed within your facility 5. Inform Stryker if any of the subject devices have been distributed to other organizations. a) Please provide contact details so that Stryker can inform the recipients appropriately. b) If you are a Distributor, note that you are responsible for notifying your affected customers. 6. Branches/Agencies: Return all affected products available at your location to: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 PR 1774317 Event 6020 7. Hospitals/Branches/Agencies: Complete and sign the enclosed Business Reply Form and fax to 1-888-230-2125 or to Email: [email protected]
Recall details
- Recall number
- Z-2407-2018
- Status
- Terminated
- Initiated by the company
- Apr 27, 2018
- Recalling firm
- Stryker GmbH
- Product code
- HSB
- FDA submissions
- K003018
- Distribution
- The products were distributed to the following US states: AL, Co, FL, KY, NC, TN, UT, and WA.
Product description
Ankle Arthrodesis Nail, right T2 Ankle 10x300mm, Product Number 18191030S
Code information (models, lots, serial numbers)
GTIN: 04546540673398; Lot code: K078C7E
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- T2 1819-1030SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 80131).
- Z-2393-2018 · TerminatedTibial Nail, Standard T2 Tibia 12x315 mm, Product Number 18221231S
- Z-2394-2018 · TerminatedFemoral Nail, A/R T2 Femur 10x320 mm, Product Number 18251032S
- Z-2395-2018 · TerminatedFemoral Nail, A/R T2 Femur 10x340 mm, Product Number 18251034S
- Z-2396-2018 · TerminatedFemoral Nail, A/R T2 Femur 12x360 mm, Product Number 18251236S
- Z-2397-2018 · TerminatedFemoral Nail, A/R T2 Femur 15x240 mm, Product Number 18251524S
- Z-2398-2018 · TerminatedFemoral Nail, A/R, R1500 T2 Femur 10x320 mm, Product Number 18281032S
- Z-2399-2018 · TerminatedFemoral Nail, A/R, R1500 T2 Femur 12x300 mm, Product Number 18281230S
- Z-2400-2018 · TerminatedFemoral Nail, A/R, R1500 T2 Femur 12x360 mm, Product Number 18281236S
- Z-2401-2018 · TerminatedFemoral Nail, A/R, R1500 T2 Femur 12x380 mm, Product Number 18281238S
- Z-2402-2018 · TerminatedFemoral Nail, LEFT T2 GTN 10x360 mm, Product Number 18501036S
- Z-2403-2018 · TerminatedFemoral Nail, LEFT T2 GTN 12x360 mm, Product Number 18501236S
- Z-2404-2018 · TerminatedFemoral Nail, LEFT T2 GTN 14x300 mm, Product Number 18501430S
- Z-2405-2018 · TerminatedFemoral Nail, RIGHT T2 GTN 9x320 mm, Product Number 18510932S
- Z-2406-2018 · TerminatedFemoral Nail, RIGHT T2 GTN 14x460 mm, Product Number 18511446S
- Z-2408-2018 · TerminatedAnkle Arthrodesis Nail, right T2 Ankle 11x300mm, Product Number 18191130S
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page