FACS recallZ-2598-2018 · Becton, Dickinson and Company, BD Biosciences · initiated May 16, 2018

Class III Terminated All recalls posted in July 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton, Dickinson and Company, BD Biosciences or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The firm, BD Life Sciences, produced the "URGENT PRODUCT RECALL-CORRECTION" letter dated 4/6/2018 to its customers during an on-site visit. by the field service engineer made beginning 5/16/2018 to correct the issue. The visits were completed 6/15/2018. If you require further assistance, contact BD Customer/Technical Support BD Customer Support at 877.232.8995 (prompt 2) for instruments in the United States. For customers outside the US, contact your local BD representative or distributor.

Recall details

Recall number
Z-2598-2018
Recall class
Class III
Status
Terminated
Initiated by the company
May 16, 2018
Posted by the FDA
Jul 13, 2018
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Quantity
8 devices
Product code
OYE
FDA submissions
K170974
Distribution
US Distribution was made to FL, KS, KY, MA, NJ, and SD.

Product description

BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only. The firm name on the label is Becton, Dickinson and Company, BD Biosciences, San Jose, CA. The Loader offers various settings to resuspend and mix samples. It can draw from 12 x 75-mm tubes in 30- and 40-tube racks. A barcode reader verifies the ID on tube racks and individual tubes in 30- tube racks. A built-in imaging system provides safety checks, such as verifying the correct rack type and tube layout, and ensures the tubes were loaded correctly.

Code information (models, lots, serial numbers)

Serial #2662879000001, UDI (01)00382906628794(11)170915(21)Z662879000001;   Serial #2662879000002, UDI (01)00382906628794(11)170915(21)Z662879000002;   Serial #2662879000003, UDI (01)00382906628794(11)170915(21)Z662879000003;   Serial #2662879000004, UDI (01)00382906628794(11)170915(21)Z662879000004;   Serial #2662879000009, UDI (01)00382906628794(11)170915(21)Z662879000009;   Serial #2662879000010; UDI (01)00382906628794(11 )170915(21)Z662879000010;   Serial #2662879000011; UDI (01)00382906628794(11)170915(21)Z662879000011; and   Serial #2662879000012, UDI (01)00382906628794(11)170915(21)Z662879000012.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page