EZ1 DSP Virus Kit (48), IVD, REF 62724 intended for in… recallZ-2596-2018 · Qiagen · initiated May 28, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Qiagen or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Underfilled reagent well that could lead to incorrect results

FDA root cause category: Under Investigation by firm

What the company is doing

On May 28, 2018 QIAGEN issued an Urgent Notice of Product Field Action and a Acknowledgement of Receipt form to customers. Actions to be taken by the customer/user/commercial partner are as follows: - If you have any remaining stocks, do not use them, contact our technical service for a free-of-charge replacement and dispose of the product in accordance with your national, local safety and environmental regulations. - Review the notice with your laboratory/medical director. - Forward the information to all individuals and departments within your organization. - If product was further distributed, forward a copy of the notice to the product end user. - QIAGEN also recommends a review of the corresponding results. Consider the possibility of under-quantification and false negative results and evaluate the need for retesting with different EZ1 DSP Virus Kit lots e.g. depending on the clinical presentation and history of the patient - Complete and return the Acknowledgement of Receipt form and email to: [email protected].

Recall details

Recall number
Z-2596-2018
Status
Terminated
Initiated by the company
May 28, 2018
Recalling firm
Qiagen
Quantity
287 pcs
Product code
JJH
Distribution
Domestic - MD Internationally - AE, AU , CZ, DE, DZ, ES, FR, GB, HK , IE, IT, NO, RO & SA

Product description

EZ1 DSP Virus Kit (48), IVD, REF 62724 intended for in vitro diagnostic use.

Code information (models, lots, serial numbers)

UDI - 04053228000099 Lot - 160020069 Expiration - 11/01/19