bioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2… recallZ-2817-2018 · bioMerieux, Inc. · initiated Jan 26, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from bioMerieux, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product.

FDA root cause category: Under Investigation by firm

What the company is doing

bioMerieux sent an Urgent Product Correction Notice letter to their French customers on January 26, 2018. Customer letters in the Americas were sent August 21, 2018. The letters identified the affected product, problem, and actions to be taken. Customers should follow steps outlined in the letter and contact bioMerieux for any additional questions regarding these corrective actions. Note: the actions do not include removal of the product. For any questions or concerns, contact your local bioMerieux representative.

Recall details

Recall number
Z-2817-2018
Status
Terminated
Initiated by the company
Jan 26, 2018
Posted by the FDA
Aug 30, 2018
Recalling firm
bioMerieux, Inc.
Quantity
13,961,320 US
Product code
LON
FDA submissions
K053097
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Bermuda, Brazil, Canada, China, Curacao, Costa Rica, Dominican Republic, Ecuador, France, Guatemala, Indonesia, Israel, Hong Kong, South Korea, South Africa, Mexico, Myanmar, Nicaragua, Peru, Philippines, Paraguay, Singapore, Thailand, El Salvador, and Vietnam.

Product description

bioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMS General Susceptibility Card Cefotoxin and AST for Oxacillin bioMerieux VITEK 2 AST-P631 bioMerieux VITEK 2 AST-GP67 Test Kit bioMerieux VITEK 2 AST-GP71 Test Kit bioMerieux VITEK 2 AST-GP75 Test Kit bioMerieux VITEK 2 AST-GP78 Test Kit Components for an automated invitro diagnostic system consisting of instruments, software and reagent cards designed for the ID and antimicrobial susceptibility testing of bacteria and yeast.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

VITEK 2 AST-P631, Ref # 414961 UDI 03573026426804 OXSF, multiple Gram Positive Cards all lots of cards with Oxacillin and Cefotoxin   OX, OXSF, multiple Gram Positive Cards  all lots of cards with Oxacillin and Cefotoxin  VITEK 2 AST-GP67 Test Kit, Ref: # 22226 UDI 03573026242060, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP71 Test Kit, Ref: #410750 UDI 03573026355173, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP75 Test Kit, Ref: #415670 UDI 03573026437183, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP78 Test Kit, Ref: #421051 UDI 03573026553982, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin.