Skip to content

DORO LUCENT Base Unit, REF 1101.021 Headrest System used… recallZ-2970-2018 · Pro-Med Instruments Gmbh · initiated Jul 5, 2018

Class I Terminated All recalls posted in September 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Pro-Med Instruments Gmbh or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.

FDA root cause category: Component change control

What the company is doing

The firm initiated the recall by email on 07/05/2018. This stage is only relevant for hospitals / users who are in possession with concerned products of two different revisions (mix-up): a) Discontinue the use of the combination and quarantine the concerned products b) Contact pro med instruments for further instructions relating to temporary replacement possibilities

Recall details

Recall number
Z-2970-2018
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Jul 5, 2018
Posted by the FDA
Sep 17, 2018
Recalling firm
Pro-Med Instruments Gmbh
Quantity
30 units
Product code
HBL
FDA submissions
K032331
Distribution
Worldwide distribution to US states of CA, TX, TN, MS, GA, WV, and OH, Qatar, India, UK, Singapore, Australia, Germany, South Africa, Italy, and Japan

Product description

DORO LUCENT Base Unit, REF 1101.021 Headrest System used in open and percutaneous craniotomies and spinal surgeries

Code information (models, lots, serial numbers)

Serial Numbers: 1704002 1704004 1704005 1704007 1704011 1704041 1704010 1704019 1704012 1704014 1704016 1704017 1704021 1704024 1704025 1704036 1704038 1704018 1704008 1704034 1704026 1704027 1704028 1704029 1704030 1704032 1704035 1704006 1704037 1704039

Other devices under the same FDA submission

1 device cleared under K032331 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 80628).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page