Philips SureSigns VSV (Vital Signs Viewer), 863067 recallZ-2892-2018 · Philips Electronics North America Corporation · initiated Jul 13, 2018
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Electronics North America Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
FDA root cause category: Labeling design
What the company is doing
On July 13, 2018, the firm issued an "URGENT-Medical Device Correction" letter dated July 10, 2018, to affected customers to advise them that that monitors with batteries that have exceeded their specified replacement interval may overheat or ignite. The firm issued a Service Guide addendum included with the letter, which contains information to assist with managing the battery and its replacement. Customers were advised to do the following: Upon receipt of this notification, carefully read the enclosed SureSigns Addendum to the Service Guide. VS Series Customers -Addendum to the Service Guide: 453564807541 VM Series Customers -Addendum to the Service Guide: 453564807551 VSV Customers- Addendum to the Service Guide: 453564815301 Promptly perform the Battery Maintenance that is called out in the Addendum for each of your affected Philips SureSigns VS & VM Monitors and View Station (VSV) by determining if the Battery Information Screen indicates if the battery cycle count exceeds the limit of 300 cycles, or the battery is older than three (3) years. If so, the battery needs to be replaced. The replacement battery can be ordered using the standard Philips replacement processes. Details on how to replace the battery can be found in the Philips SureSigns VS & VM Monitors and View Station (VSV) Service Guide. Review this information with all staff members who are responsible for device management of the Philips SureSigns VS & VM Monitors and View Station (VSV). Please store the below Service Guide Addendum with your Philips SureSigns VS & VM Monitors and View Station (VSV) Service Guide documentation. VS Series Customers -Addendum to the Service Guide: 453564807541 VM Series Customers -Addendum to the Service Guide: 453564807551 VSV Customers- Addendum to the Service Guide: 453564815301 Complete and return the response card provided. In addition to providing the Service Guide Addendum and this notice, Philips plans to release a customer ins
Recall details
- Recall number
- Z-2892-2018
- Status
- Open, Classified
- Initiated by the company
- Jul 13, 2018
- Recalling firm
- Philips Electronics North America Corporation
- Quantity
- VM 96,324 Units; VS 110,424 Units; VSV 691 Units
- Product code
- MHX
- FDA submissions
- K070588
- Distribution
- Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic, Dominican Republic, El Salvador, Egypt, Estonia, Ethiopia, France, French Guiana, French Polynesia, Finland, Guadeloupe, Germany, Ghana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Kenya, Korea, Kosovo, Republic of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Lybia, Macao, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, Zambia and Zimbabwe.
Product description
Philips SureSigns VSV (Vital Signs Viewer), 863067 Indicated for central monitoring of multiple adult, pediatric and neonatal patients, and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric and neonatal patients to gain information for the treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.
Code information (models, lots, serial numbers)
UDI# VSV (863067): 00884838009578
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SureSigns VSV Vital Signs ViewStationPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 80645).
- Z-2894-2018 · Open, ClassifiedPhilips SureSigns VS4 Vital Signs Monitor, 863283 863286 The SureSigns VS4 vital signs monitor is for use by healthcare professionals whenever there is a need for monitoring the …
- Z-2895-2018 · Open, ClassifiedPhilips SureSigns VM 3/4/6/8 Patient Monitors, 863063 863064 863065 863066 863068 863077 863085 863086 863317 863287 863288 863289 The SureSigns VM4, VM6, and VM8 patient monitor…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.