Diagnostica Stago STA Noplastine CI Plus, 10 ml vial.… recallZ-2993-2018 · Diagnostica Stago, Inc. · initiated Jul 24, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

QC values outside of the assigned ranges (prolonged Prothrombin time / decreased PT %)

FDA root cause category: Device Design

What the company is doing

Diagnostica Stago issued notification on July 24, 2018 to US Customers provided the reason for the notification, actions to be taken to mitigate the risk. Action: If you have in your laboratory, any of the lots among those listed in the appendix, we are asking you to: If it is not already done, run a Quality Control test at every change of vial. Return to your Stago affiliate, by fax or e-mail, the enclosed form completed and confirming that you have read this letter. Questions contact Stago Hotline (1-800-725-0607)

Recall details

Recall number
Z-2993-2018
Status
Terminated
Initiated by the company
Jul 24, 2018
Recalling firm
Diagnostica Stago, Inc.
Quantity
24043 kits
Product code
GJS
FDA submissions
K922040
Distribution
Nationwide

Product description

Diagnostica Stago STA Noplastine CI Plus, 10 ml vial. In-Vitro Diagnostic for determination of the prothrombin time (PT) in plasma. (REF 00667) The STA - Noplastine CI Plus kits provide reagents for the determination of the prothrombin time (PT) in plasma with STA-R, STA Compact and STA Satellite. (In the USA this procedure has been assigned to the moderate complexity category per CLIA 1988 - CDC Analyte Code 4922; CDC Test System Codes 4677 and 4875).

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Lot Number/Exp. Date/UDI 250791 2018-07-31  (01)03607450006674(11)160731(17)180731(10)250791(241)00667; 250792 2018-07-31 (01)03607450006674(11)160731(17)180731(10)250792(241)00667; 250911 2018-08-31 (01)03607450006674(11)160831(17)180831(10)250911(241)00667; 251231 2018-10-31  (01)03607450006674(11)161031(17)181031(10)251231(241)00667; 251311 2018-11-30 (01)03607450006674(11)161130(17)181130(10)251311(241)00667; 251364 2018-11-30 (01)03607450006674(11)161130(17)181130(10)251364(241)00667; 251427 2018-11-30  (01)03607450006674(11)161130(17)181130(10)251427(241)00667; 251528 2018-12-31  (01)03607450006674(11)161231(17)181231(10)251528(241)00667; 251707 2019-01-31  (01)03607450006674(11)170131(17)190131(10)251707(241)00667; 251708 2019-01-31  (01)03607450006674(11)170131(17)190131(10)251708(241)00667; 251767 2019-02-28  (01)03607450006674(11)170228(17)190228(10)251767(241)00667; 251800 2019-02-28  (01)03607450006674(11)170228(17)190228(10)251800(241)00667; 252534 2019-07-31  (01)03607450006674(11)170731(17)190731(10)252534(241)00667; 252737 2019-09-30  (01)03607450006674(11)170930(17)190930(10)252737(241)00667

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 80826).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page