Accu-Temp recallZ-3045-2018 · Beaver Visitec · initiated Jul 19, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beaver Visitec or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A small number of devices in this lot may have a melted cap and a hole formed through the packaging. The risk is potentially the product will no longer be sterile for use and may cause harm to a patient.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Urgent Medical Device Recall notification letters dated 7/19/18 were distributed to customers. The letters instructed customers to perform the following actions: 1. Identify the product. 2. Immediately examine your inventory and quarantine product subject to recall. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. 3. Consider all potential users of this product in your wholesale and/or user level supply chain. Your assistance is appreciated and necessary to prevent patient harm. 4. Follow the attached Destruction Instructions to disable the wire tip to safely destroy this product to prevent activation or future use of this product. 5. Complete and return the enclosed response form as soon as possible. Please provide objective evidence of the product destruction and we will credit your account. Customers were also contacted via phone beginning 8/1/18

Recall details

Recall number
Z-3045-2018
Status
Terminated
Initiated by the company
Jul 19, 2018
Recalling firm
Beaver Visitec
Quantity
800 units
Product code
HQP
FDA submissions
K771189
Distribution
Worldwide Distribution - US Nationwide in the states of: CA, DC, FL, IA, ID, IL, MA, MI, MN, NY, OK, PA, TN, TX, UT, and WA and the countries of Australia, France, Netherlands, Spain, and United Kingdom.

Product description

Beaver Accu-Temp High Temperature Cautery, 2 inch Shaft with Fine Tip, Model Number 8445000 Product Usage: The indications and intended use of the High Temp cautery are for coagulating tissue or arresting bleeding from small vessels by heat conducted through the wire tip. Indications include ophthalmic, general and plastic surgery.

Code information (models, lots, serial numbers)

Lot 6002809, UDI 8445000 30886158010297