Skip to content

Cypass recallZ-0185-2019 · Alcon Research, LTD. · initiated Aug 28, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Alcon Research, LTD. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.

FDA root cause category: Device Design

What the company is doing

The firm initiated their market withdrawal by letter on 08/28/2018. The firm requested surgeons immediately cease implantation of the device and return any unused product.

Recall details

Recall number
Z-0185-2019
Status
Terminated
Initiated by the company
Aug 28, 2018
Posted by the FDA
Oct 23, 2018
Recalling firm
Alcon Research, LTD.
Quantity
30561 units
Product code
OGO
FDA submissions
P150037
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.

Product description

CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use.

Code information (models, lots, serial numbers)

GTIN: 00380657540037 All lot/serial numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 80931).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page