Skip to content

Reprocessed Agilis Steerable Introducer: indicated for… recallZ-0090-2019 · Sterilmed, Inc. · initiated Sep 17, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sterilmed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Reprocessed Agilis Steerable Introducers may not meet endotoxin requirements per product specification for biological residues determined by LAL testing.

FDA root cause category: Material/Component Contamination

What the company is doing

Sterilmed notified customers on about 09/17/2018, via URGENT MEDICAL DEVICE RECALL letter sent via UPS next day. Instructions included examine inventory for affected devices, remove and return affected devices to Stericycle (3rd party processor), provide recall information to appropriate personnel at the facility, and complete and return the Business Reply Form to Stericycle. For questions, contact Sterilmed's Customer Care Team at (888) 541-0078, Monday through Friday from 7:00 AM to 5:00 PM CST.

Recall details

Recall number
Z-0090-2019
Status
Terminated
Initiated by the company
Sep 17, 2018
Posted by the FDA
Oct 9, 2018
Recalling firm
Sterilmed, Inc.
Quantity
218 units
Product code
PNH
FDA submissions
K152090
Distribution
U.S. Nationwide distribution in the states of CA, CO, MN, and WA.

Product description

Reprocessed Agilis Steerable Introducer: indicated for introducing various cardiovascular catheters into the heart. (a) small curl (16.8mm curve type/71 cm length), Product Code STJ408309; (b) medium curl (22.4 mm curve type/71 cm length), Product Code STJ408310; (c) large curl (50 mm curve type/71 cm length), Product Code STJ408324

Code information (models, lots, serial numbers)

All lots with expiration dates prior to 2019-06-08 of the following: (a) Product Code STJ408309, GTIN 10888551044912  (b) Product Code STJ408310, GTIN 10888551044929 (c) Product Code STJ408324, GTIN 10888551044950

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page