BD Nexiva, BD Q-Syte recallZ-0384-2019 · Becton Dickinson & Company · initiated Aug 23, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
One lot of the specified needles is not performing as intended due to a failure in the needle tip shield/safety mechanism, which may lead to an increased risk of a contaminated needlestick injury.
FDA root cause category: Under Investigation by firm
What the company is doing
Urgent Medical Device Recall notification letters dated 8/23/18 were sent to customers. The letters instruct customers to perform the following actions: 1. Immediately review your inventory for the specific Catalogue (Ref) and lot number listed above and destroy all product subject to the recall following your institutions process for destruction. 2. Share this recall notification with all users of the product to ensure that they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program. Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787 Actions Taken Distributors were instructed to: 1. Immediately review your inventory for the specific Catalogue (Ref) and lot number listed above and return all product subject to recall according to the Packing Instructions enclosed. 2. Identify all your customers that purchased any affected product, as defined in this recall notification. E-mail an excel file listing of all customers to [email protected], within 72 hours of receipt of this letter. 3. Complete the attached Wholesaler/Distributor Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your credit accordingly. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program. Web: MedWatch website at www.fda.gov/medwatch Phone: 1
Recall details
- Recall number
- Z-0384-2019
- Status
- Terminated
- Initiated by the company
- Aug 23, 2018
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 133,230 units
- Product code
- FOZ
- FDA submissions
- K032843
- Distribution
- Worldwide Distribution -US Nationwide in the states of: CA, CO, FL, GA, IL, LA, MI, ND, NE, NY, OH, PA, RI, SC, VA, WA and WV. The products were distributed to the following foreign countries: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Costa Rica, Mexico and New Zealand.
Product description
BD NEXIVA Closed IV Catheter System Dual Port 18GA 1.25 IN (BD Nexiva Catheter), Catalog Number 383539 Product Usage: BD Nexiva Closed IV Catheter System Dual Port (hereafter referred to as BD Nexiva device or catheter) (see Figure below) incorporates a passive needle shielding mechanism that is activated as part of the usual insertion procedure when the needle is withdrawn from the catheter. BD Nexiva device is designed to protect the clinician from accidental needle stick injuries and minimize exposure to blood during the insertion process. The needle shield is designed to not release from the catheter adapter until the needle tip is protected and cannot be rearmed once activated. The BD Nexiva device is intended for use in the peripheral vascular system to provide access for the administration of fluids, medication, blood and blood products, and for the withdrawal of blood samples. This catheter may be used for any patient population with consideration given to patient size, appropriateness for solutions being infused and duration of therapy.
Code information (models, lots, serial numbers)
Lot 8087800, UDI (01)30382903835394
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BD Nexiva, BD Q-Syte 383539Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall170 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page