Trinias recallZ-0197-2019 · Shimadzu Medical Systems Usa Com · initiated Jun 15, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Shimadzu Medical Systems Usa Com or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a software issue the C-arm may unexpectedly moved when being moved to the park position and during specific operations. There is a possibility the C-arm will come into contact with a patient or healthcare personnel.
FDA root cause category: Software design
What the company is doing
On June 15, 2018, the firm sent an URGENT Voluntary Medical Device Recall Notice to its customer via certified mail informing the customer that the Trinias C-arm could unexpectedly move while being moved to the park position and/or when performing specific operations. An investigation has shown that a software defect in the MH300, which is a floormount C-arm tube support of the Trinias biplane system, caused this event under specific conditions described as below. Neither the Trinias single plane system nor the Trinias dual plane system have the software defect. If the incident were to recur, there is a possibility the C-arm will come into contact with a patient or healthcare personnel. The specific conditions are as follows: (1) The table top is positioned at head side so that the C-arm safety interlock is active. (2) If operator performs a parking operation and presses a direct memory switch at the same time, the the software runs abnormally, and the C-arm does not movebecause the safety interlock is active (1). (3) If the operator then moves the table top to foot side, the C-arm safety interlock is cancelled. (4) While the C-arm safety interlock is cancelled due to the actions performed in (3), the C-arm starts moving without the operators control. Therefore, the firm will implement a voluntary recall/field corrective action to revise the software, thereby fixing the problem. This Field Safety Notice is intended to inform the customer about: - the description of the issue - the actions planned by Shimadzu to correct the issue - the actions that should be taken by the customer How to identify the affected products: The Shimadzu Digital Angiographic System Trinias; Serial Number 41E58C977001 Please contact the local Shimadzu representative or contact Shimadzu National Technical Support. Corrective action plan by Shimadzu: After the software defect has been corrected and the correction safety and efficacy have been verified through our verifi
Recall details
- Recall number
- Z-0197-2019
- Status
- Terminated
- Initiated by the company
- Jun 15, 2018
- Posted by the FDA
- Oct 19, 2018
- Recalling firm
- Shimadzu Medical Systems Usa Com
- Quantity
- 1
- Product code
- OWB
- FDA submissions
- K123508
- Distribution
- Domestic: LA
Product description
TRINIAS Digital Angiographic System intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures.
Code information (models, lots, serial numbers)
Model Trinias UDI (01)04540217049080(11)170727(21)41E58C977001 Serial Number 41E58C977001
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Trinias 503-79402-33Shimadzu Corporation · System, X-Ray, Fluoroscopic, Image-Intensified1 recall0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K123508 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Trinias 503-79402-31Shimadzu Corporation · System, X-Ray, Fluoroscopic, Image-Intensified1 recall0 reports
- Trinias 503-79830-31Shimadzu Corporation · System, X-Ray, Fluoroscopic, Image-Intensified1 recall0 reports
- Trinias 503-79821-31Shimadzu Corporation · System, X-Ray, Fluoroscopic, Image-Intensified1 recall0 reports
- Trinias 503-79822-31Shimadzu Corporation · System, X-Ray, Fluoroscopic, Image-Intensified1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page