SYNCHRO 2 recallZ-0227-2020 · Stryker Neurovascular · initiated Jun 8, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Neurovascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The FDA record does not include a reason.
FDA root cause category: Employee error
What the company is doing
On 6/08/18, "Urgent Medical Device Voluntary Customer Communication" were mailed via courier. The firm informed its customers that it has become aware that the product label (pouch and carton) for the affected products contain an incorrect value in millimeters (0.036 mm) for the Guidewire Outer Diameter dimension, but there will be no products being recalled. The Customer Notification Letter instructs customers to read this communication carefully and complete the following actions: 1. Immediately check your inventory. 2. Review the Communication and ensure full understanding of the contents. 3. Circulate this Customer Communication internally to all interested/affected parites. 4. Inform the Recalling Firm if any of the subject devices have been distributed to other organizations. 5. Please provide contact details so that the Recalling Firm can inform the recipients appropriately. 6. Please inform the Recalling Firm of any adverse events concening the use of the subject devices. 7. Return the completed form to your nominated Sales Representative of to [email protected]. 8. The Recalling Firm is requesting that customers respond to the Customer Communication within 7 calendar days for the date of receipt and a target date for completion of this action is 30 June 2018.
Recall details
- Recall number
- Z-0227-2020
- Status
- Terminated
- Initiated by the company
- Jun 8, 2018
- Posted by the FDA
- Oct 29, 2019
- Recalling firm
- Stryker Neurovascular
- Quantity
- 831
- Product code
- DQX
- FDA submissions
- K190843
- Distribution
- Worldwide distributions- U.S. Nationwide including the states of AL, AZ, AR, CA, CO, CT, DC, DE, FL, GA, GU, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY. Countries of Austria, Argentina, Australia, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Lithuania, Mexico, Netherlands, Norway, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, UAE, and UK.
Product description
Synchro2, 0.014in x 300cm SOFT Guidewire with Hydrophilic Coating , UPN Product No. M00326310, REF 2631, STERILE:EO - Product Usage: product is intended for general intravascular use, including neurovascular and peripheral vasculatures.
Code information (models, lots, serial numbers)
UPN: M00326310 Lot Numbers/UDI: 0000000563/ (01)04546540688743(17)201031(10)0000000563; 0000000986/ (01)04546540688743(17)201031(10)0000000986; 0000001646/ (01)04546540688743(17)201130(10)0000001646; 0000002238/ (01)04546540688743(17)201130(10)0000002238; 0000002426/ (01)04546540688743(17)201231(10)0000002426; 0000002640/ (01)04546540688743(17)201231(10)0000002640; 0000003465/ (01)04546540688743(17)210131(10)0000003465; 0000004906/ (01)04546540688743(17)210228(10)0000004906; 0000005147/ (01)04546540688743(17)210228(10)0000005147; 0000005343/ (01)04546540688743(17)210228(10)0000005343; 0000007637/ (01)04546540688743(17)210421(10)0000007637;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SYNCHRO 2 M00326310Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls5 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 81108).
- Z-0225-2020 · TerminatedSynchro2, 0.014 in x 200cm, SOFT Guidewire with Hydrophilic Coating, UPN Product No. M00326010 REF 2601, STERILE: EO - Product Usage: product is intended for general intravascular …
- Z-0226-2020 · TerminatedSynchro2, 0.014in x 200cm, SOFT PRE-SHAPED Guidewire with Hydrophilic Coating, UPN Product No. M00326110, REF 2611, STERILE: EO - Product Usage: product is intended for general in…
- Z-0228-2020 · TerminatedSynchro2, 0.014in. x 300cm, SOFT PRE-SHAPED Guidewire with Hydrophilic Coating, UPN Product No. M00326320, REF 2632, STERILE:EO - Product Usage: product is intended for general int…
- Z-0229-2020 · TerminatedSynchro2, 0.014in x 200cm STANDARD Guidewire with Hydrophilic Coating, UPN Product No. M00326410, REF 2641, STERILE: EO - Product Usage: product is intended for general intravascul…
- Z-0230-2020 · TerminatedSynchro2, 0.014in. x 200cm STANDARD PRE-SHAPED Guidewire with Hydrophilic Coating UPN Product No. M00326420, REF 2642, STERILE: EO - Product Usage: product is intended for general …
- Z-0231-2020 · TerminatedSynchro2, 0.014in x 300cm STANDARD Guidewire with Hydrophilic Coating UPN Product No. M00326510, REF 2651, STERILE:EO - Product Usage: product is intended for general intravascular…
- Z-0232-2020 · TerminatedSynchro2, 0.014in x 300cm STANDARD PRE-SHAPED Guidewire with Hydrophilic Coating, UPN Product No. M00326520, REF 2652, STERILE: EO - Product Usage: product is intended for general …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page