Luminex Aries M1 System recallZ-2292-2019 · Luminex Corporation · initiated Jun 29, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Luminex Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient order.
FDA root cause category: Process control
What the company is doing
On July 10, 2018, the firm sent an "URGENT: FIELD SAFETY NOTICE" to customers via email (with a follow-up telephone call) informing them that they may have received an affected barcode scanner. The Notice explained that when the potentially affected barcode scanner is used on multiple ARIES cassettes at the same time, the system may prepopulate the Sample ID field with a previously stored sample ID. This field will be overwritten by scanning or manually entering a sample, but in the event the ARIES cassette is not scanned, the prepopulated field may cause the incorrect association of sample identification to occur. The firm has conducted internal investigation that have shown that the firmware version on hand-held barcode scanner contained the incorrect version of firmware approved for use with the ARIES Systems. Given that the ARIES System barcode scanner (CN-0321-01 /JDK-2330 Rev B) has the potential to contribute to an incorrect result by creating the possibility that a Patient/Sample ID may be assigned to the wrong cassette, we are voluntarily initiating a recall to ensure no additional handheld scanners have this discrepant firmware. There have not been any adverse events associated with the use of the ARIES System barcode scanners at this time. To confirm whether or not the customer's particular barcode scanner is affected, the firm is advising its customer to verify the firmware version of the handheld barcode scanner by following the attached instructions ("Confirming the Firmware Version on the ARIES Hand-Held Barcode Scanner") . An affected unit will include the labeling JDK-2330 Rev B on the underside of the scanner AND will have firmware version PAACES00-002-R03 or CAACES00-002-R03. Please contact Technical Support if you identify a discrepant barcode scanner model. Correct Model and Firmware (below) Model/Rev. Firmware Version JDK-2330 Rev B. PAACES00-001-R03 CAACES00-001-R03 If the in
Recall details
- Recall number
- Z-2292-2019
- Status
- Terminated
- Initiated by the company
- Jun 29, 2018
- Posted by the FDA
- Aug 19, 2019
- Recalling firm
- Luminex Corporation
- Quantity
- 153 (140 U.S. and 13 O.U.S.)
- Product code
- OOI
- FDA submissions
- K151917, K160517, K161495, K162772
- Distribution
- Worldwide distribution. US nationwide, Canada, Ireland, Netherlands, Singapore, Thailand, and United Kingdom
Product description
Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical laboratories. Scanner Label/Labeling: JADAK JDK-2330 REV B 170512-017
Code information (models, lots, serial numbers)
ARIES System UDI: 00840487101537 ARIES M1 System UDI: 00840487100080 ARIES-M12V1-IVD, ARIES System ARIES-M6V1-IVD, ARIES M1 System CN-0321-01, ARIES System Handheld Barcode Scanner All scanners manufactured after April 2017 and including Part Number JDK-2330 Rev B.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ARIES® M1 System ARIES M6V1Luminex Corporation · real time Nucleic acid amplification system2 recalls0 reports
- ARIES® System ARIES M12V1Luminex Corporation · real time Nucleic acid amplification system2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page