Skip to content

VITROS recallZ-0653-2019 · Ortho-Clinical Diagnostics · initiated Oct 19, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.

FDA root cause category: Software Manufacturing/Software Deployment

What the company is doing

On October 19, 2018, Ortho Clinical Diagnostics issued "URGENT PRODUCT CORRECTION NOTIFICATION" notices via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers to inform them of the issue and actions to take to prevent occurrence of the anomaly. On November 19, 2018 Ortho Clinical Diagnostics re-issued correction notices to customers pertaining to the 5,1 FS system (CL2018-194_5,1FS). Customers may continue using their VITROS Systems and have been informed that the software issue will be mitigated with the next version of software. The expected dates of these software versions are included in the communication. Foreign affiliates were informed by email approximately on 21 October 2018 of the issue and instructed to notify their consignees of the issue and required actions to prevent the occurrence of the anomaly. Please refer to the Customer letter (Ref. CL2018-194_51FS, CL2018-194_3600, CL2018-194_4600, and CL2018-194_5600) for specific details on Required Actions by User. If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.

Recall details

Recall number
Z-0653-2019
Status
Terminated
Initiated by the company
Oct 19, 2018
Recalling firm
Ortho-Clinical Diagnostics
Quantity
728
Product code
JJE
FDA submissions
K063144
Distribution
US Nationwide Distribution

Product description

VITROS 4600 Chemistry System, Version 3.3.1 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the analyte) such as serum, plasma, urine and cerebrospinal fluid.

Code information (models, lots, serial numbers)

Product Code: 6802445 UDI: 10758750012343 Serial Numbers: J46000108-J46001132

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 81563).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page