Discovery recallZ-0606-2019 · GE Healthcare, LLC · initiated May 24, 2018

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.

FDA root cause category: Component change control

What the company is doing

The firm initiated the recall by letter on 05/24/2018. The letter explained the issue and requested the UPS(s) be disconnected from the system until the system could be repaired.

Recall details

Recall number
Z-0606-2019
Status
Completed
Initiated by the company
May 24, 2018
Recalling firm
GE Healthcare, LLC
Quantity
4 units
Product code
KPS
FDA submissions
K093514
Distribution
CA, DC, OH, UT, WI, Canada

Product description

Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.

Code information (models, lots, serial numbers)

GTIN 00840682124393 Serial Numbers: DDRW34032, DDRX34038, DDRX34041, DDRX34045

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

5 devices cleared under K093514 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 81680).