Skip to content

ARCHITECT recallZ-1296-2019 · Abbott Laboratories, Inc · initiated Aug 28, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Laboratories, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential to generate falsely elevated serum or plasma chloride results when using the product.

FDA root cause category: Under Investigation by firm

What the company is doing

The recalling firm issued letters dated 8/28/2018 via overnight mail. The letter explained the issue, patient impact, and instructed the consignee to immediately discontinue use of the product and destroy any remaining inventory of the impacted lot number. The letter was to be reviewed with the Medical Director and laboratory protocol should be followed regarding the need to review previously-reported patient results. A Customer Reply Form was enclosed for the U.S. consignees to indicate the consignee understood and completed the necessary actions to be taken. For questions regarding this information, U.S. Customers please contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.

Recall details

Recall number
Z-1296-2019
Status
Terminated
Initiated by the company
Aug 28, 2018
Recalling firm
Abbott Laboratories, Inc
Quantity
99 modules/kits
Product code
CGZ
FDA submissions
K980367
Distribution
Worldwide Distribution - US Nationwide in the states of AR, CA, CT, FL, IA, IN, LA, MA, MD, MS, NJ, NY, PA, SC, TX, VT, WA, and WV. There was government distribution and no military distribution. Foreign distribution was made to Australia, China, Hong Kong, and New Zealand.

Product description

Abbott ARCHITECT cSystems ICT Module, REF 09D28-03. Product Usage: The ARCHITECT cSystems ICT (Integrated Chip Technology) is used for the quantitation of sodium, potassium, and chloride in human serum, plasma, or urine.

Code information (models, lots, serial numbers)

Lot No. 180326, Exp. 12/26/2018;  UDI (01)00380740019327(17)181226(10)180326(21)180326301 through (01)00380740019327(17)181226(10)180326(21)180326399

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page